FDA Adverse Event Malfunction Summary report: N

HIP END EFFECTOR, VARIABLE ANGLE

MDR report key: 6521715 · Received April 26, 2017

Report

Report Number
3005985723-2017-00192
Event Type
Malfunction
Date Received
April 26, 2017
Date of Event
April 8, 2017
Report Date
May 19, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K141989
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: PACK SLEEVE FELL OFF THE END EFFECTOR. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION CONFIRMS FAILED LOCKING MECHANISM WITH BALL BEARINGS FAILING. ALSO VISUAL INSPECTION CONFIRMS A SHEARED OFF 202866 HIP RELEASE KNOB. REFERENCE ATTACHED IMAGES. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 08/30/2016 INCLUDING SERIAL NUMBER (B)(4). COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 206967, LOT 19410515 SHOWS 3 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THESE COMPLAINTS ARE: (B)(4). CONCLUSIONS: ALLEGED FAILURE CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA (B)(4) WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE SURGEON WAS COMPLETING A TOTAL HIP ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM. THE PACK SLEEVE FELL OFF THE END EFFECTOR.

Description of Event or Problem · 1

THE SURGEON WAS COMPLETING A TOTAL HIP ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM. THE PACK SLEEVE FELL OFF THE END EFFECTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303544 HIP END EFFECTOR, VARIABLE ANGLE STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP.

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization