FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 6519698 · Received April 25, 2017

Report

Report Number
3004209178-2017-47473
Event Type
Malfunction
Date Received
April 25, 2017
Date of Event
July 11, 2013
Report Date
July 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARMED DURING REWIND DUE TO CORRODED ON MOTOR HOME SWITCH. ALSO. CORROSION FOUND ON LCD BOARD. HD (B)(6) 2013. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A MOTOR ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS UNKNOWN. TROUBLESHOOTING WAS NOT PERFORMED. THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301193 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWL

Patients

Seq Age Sex Outcome Treatment
1