TIBIAL INLAY IMPACTOR HEAD
Report
- Report Number
- 3005738134-2017-00004
- Event Type
- Malfunction
- Date Received
- April 24, 2017
- Date of Event
- March 28, 2017
- Report Date
- July 26, 2017
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- NPJ
- PMA / PMN Number
- K090763
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
AN EVENT REGARDING CRACK/FRACTURE INVOLVING AN MAKO IMPACTOR WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: THE DEVICE WAS RETURNED IN USED CONDITION AND SHOWED THE IMPACTOR HEAD WAS FRACTURED. THERE ARE SCRATCHES AND GOUGES ON THE BOTTOM AND SIDES OF THE IMPACTOR. MATERIAL ANALYSIS INDICATED FRACTURE CONSISTENT WITH AN OVERLOAD CONDITION. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS MEDICAL RECORDS WERE NOT RECEIVED AND PATIENT FACTORS DIDN'T CONTRIBUTE TO THE EVENT. DEVICE HISTORY REVIEW: INDICATED THE DEVICES ACCEPTED INTO FINAL STOCK FROM THE REPORTED LOT WERE FREE FROM DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAS BEEN NO OTHER EVENT FOR THE LOT REFERENCED. CONCLUSIONS: THE REPORTED EVENT WAS CONFIRMED BY VISUAL INSPECTION WHICH SHOWED THE IMPACTOR HEAD FRACTURED. MATERIAL ANALYSIS INDICATED FRACTURE CONSISTENT WITH AN OVERLOAD CONDITION. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME. IF ADDITIONAL INFORMATION WILL BE RECEIVED, THIS INVESTIGATION WILL BE REOPENED AND RE-EVALUATED.
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
THE IMPACTOR TIP PART NUMBER 160177 BROKE AT THE HOSPITAL.
THE IMPACTOR TIP PART NUMBER 160177 BROKE AT THE HOSPITAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296831 | TIBIAL INLAY IMPACTOR HEAD | PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM | NPJ | MAKO SURGICAL CORP. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |