FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 6514343 · Received April 22, 2017

Report

Report Number
3004209178-2017-46324
Event Type
Malfunction
Date Received
April 22, 2017
Date of Event
March 11, 2017
Report Date
April 14, 2017
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE IDLE CURRENT MEASUREMENT TEST, RUN CURRENT MEASUREMENT TEST, OFF NO POWER TEST, A21 ERROR TEST, DISPLACEMENT TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, REWIND TEST AND EXCESSIVE NO DELIVERY TEST. HOWEVER, THE INSULIN PUMP WAS UNABLE TO VIBRATE DURING SELF TEST DUE TO FAULTY VIBRATOR MOTOR. THE INSULIN PUMP WAS RECEIVED WITH NO BATTERY INSTALLED. NO COSMETIC DAMAGES WERE NOTED. DATA ANALYSIS: (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U. (B)(6) 2017 DAILY INSULIN TOTAL = 0U.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER PASSED AWAY IN A HOSPICE HOUSE ON (B)(6) 2017. THE CAUSE OF DEATH WAS CANCER. THE CALLER STATED THAT THIS WAS THE FOURTH ROUND SINCE SEVENTEEN MONTHS. THE CALLER STATED THAT THE CUSTOMER WAS NOT WEARING THE INSULIN PUMP AT THE TIME OF DEATH; IT HAD BEEN DISCONNECTED TEN TO ELEVEN DAYS PRIOR TO PASSING BECAUSE THE HOSPITAL DID NOT KNOW HOW TO OPERATE THE INSULIN PUMP. THE CALLER DID NOT KNOW THE CUSTOMER'S BLOOD GLUCOSE AT THE TIME OF DEATH. HOWEVER, STATED THAT THE LAST KNOW BLOOD GLUCOSE VALUE WAS 121 MG/DL. THE CALLER AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS. PLEASE SEE SECTION H10 FOR ANALYSIS RESULTS.