FDA Adverse Event Malfunction Summary report: N

2517506-2017-00415

MDR report key: 6513049 · Received April 21, 2017

Report

Report Number
2517506-2017-00415
Event Type
Malfunction
Date Received
April 21, 2017
Date of Event
March 30, 2017
Report Date
April 21, 2017
PMA / PMN Number
K051087
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS HEALTHCARE DIAGNOSTICS INC. (SIEMENS) CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE CSE ALIGNED AND CLEANED REAGENT PROBES 1 AND 2 AND PRIMED REAGENT PROBES 1 AND 2 AND SAMPLE PROBE 1 (S1). SERVICE METHODS WERE PERFORMED. DURING FOLLOW-UP VISITS, THE CSE REPLACED THE ALIQUOT PROBE AND THE INTEGRATED MULTI-SENSOR TECHNOLOGY PROBE AND RAN A QUICK CHECK. THE CSE CLEANED ALL DRAINS AND ADJUSTED VACUUM, REPLACED S1 MIXER AND PROBE AND PERFORMED MIXER DIAGNOSTICS TEST. QUICK CHECK AND ALL FUNCTIONAL TESTS WERE ACCEPTABLE. THE CSE CALIBRATED AND RAN QC FOR HEMOGLOBIN A1C AND PERFORMED PRECISION TESTING ON PATIENT SAMPLES, ALL OF WHICH WERE ACCEPTABLE. THE CAUSE OF THE DISCORDANT HEMOGLOBIN A1C RESULTS IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 1

DISCORDANT HEMOGLOBIN A1C RESULTS WERE OBTAINED FROM A DIMENSION VISTA 500 INSTRUMENT. THE RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULTS. REPEAT TESTING USING THE SAME SAMPLES WAS PERFORMED ON THE SAME DIMENSION VISTA 500 INSTRUMENT. SOME SAMPLES WERE REPEATED A SECOND TIME. THE REPEAT RESULTS WERE REPORTED TO THE PHYSICIAN(S).

Patients

Seq Age Sex Outcome Treatment
1