FDA Adverse Event Malfunction Summary report: Y

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 6511334 · Received April 21, 2017

Report

Report Number
2531779-2017-08272
Event Type
Malfunction
Date Received
April 21, 2017
Report Date
March 31, 2017
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

OF THE 42 ALLEGATIONS OF A MALFUNCTION, 30 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 1 WAS FOUND TO BE MALFUNCTIONING DUE TO A CALL SERVICE ALARM ISSUE. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Description of Event or Problem · 0

THIS REPORT SUMMARIZES 42 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 49-228 POUNDS AND BETWEEN 5 AND 80 YEARS OF AGE. OF THE REPORTED PATIENTS, 18 WERE MALE, 23 WERE FEMALE, AND 1 WAS UNKNOWN.

Additional Manufacturer Narrative · 1

FOLLOW UP #1 07/29/2017. DEVICE EVALUATION: IN Q2 2017, THERE WERE 8 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER (B)(4). THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294644 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1