FDA Adverse Event Malfunction Summary report: Y

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 6511148 · Received April 21, 2017

Report

Report Number
2531779-2017-08252
Event Type
Malfunction
Date Received
April 21, 2017
Report Date
March 31, 2017
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

OF THE 6 ALLEGATIONS OF A MALFUNCTION, 3 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 2 WERE FOUND TO BE MALFUNCTIONING DUE TO A DISPLAY CLOUDY WITH MOISTURE ISSUE. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Additional Manufacturer Narrative · 0

FOLOW UP #1 07/29/2017 DEVICE EVALUATION: IN Q2 2017, THERE WERE 4 INSTANCES OF CLOUDY W/ MOISTURE FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Description of Event or Problem · 0

THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CARTRIDGE COMPARTMENT AND CAP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 183-210 POUNDS AND BETWEEN 2 AND 68 YEARS OF AGE. OF THE REPORTED PATIENTS, 3 WERE MALE AND 3 WERE FEMALE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294347 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1