COBAS E 411 IMMUNOASSAY ANALYZER
Report
- Report Number
- 1823260-2017-00834
- Event Type
- Malfunction
- Date Received
- April 20, 2017
- Date of Event
- April 4, 2017
- Report Date
- May 23, 2017
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SI
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WAS PROVIDED THAT THE PATIENT WAS NOT INCORRECTLY TREATED BASED ON THE ERRONEOUS RESULT. THE FIELD SERVICE REPRESENTATIVE FOUND FOAM IN THE REAGENT PACK SO HE ADJUSTED THE BEAD MIXER PADDLE.
(B)(4). THIS EVENT OCCURRED IN (B)(6).
CLARIFICATION WAS PROVIDED THAT THE FIELD SERVICE REPRESENTATIVE REPLACED THE BEAD MIXER PADDLE REPLACEMENT AND THEN PERFORMED ADJUSTMENTS. THE CUSTOMER STATED CALIBRATION AND QC RESULTS WERE IN THE ACCEPTABLE RANGE AFTER THE SERVICE VISIT. THE CUSTOMER HAS NOT REPORTED ANY ADDITIONAL EVENTS. THE INVESTIGATION COULD NOT DETERMINE A SPECIFIC ROOT CAUSE, BUT COULD NOT EXCLUDE THE FOAM ON THE REAGENT SURFACE HAD AN IMPACT ON THE REPORTED SAMPLE RESULT.
THE CUSTOMER RECEIVED QUESTIONABLE RESULTS FOR ELECSYS PROBNP II IMMUNOASSAY AND ELECSYS PTH IMMUNOASSAY. OF THE DATA PROVIDED, ONLY THE RESULTS FOR PTH WERE DISCREPANT AND REPORTED OUTSIDE THE LABORATORY. THE INITIAL RESULT FROM THE PRIMARY SAMPLE TUBE WAS 17.60 NG/L AND WAS REPORTED. THE REPEAT RESULTS FROM AN ALIQUOT IN A CUP WERE 1288 NG/L AND 1297 NG/L. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE REAGENT LOT NUMBER WAS 143251. THE EXPIRATION DATE WAS REQUESTED BUT WAS NOT PROVIDED. THE CUSTOMER STATED ALL CALIBRATION AND QC RESULTS WERE WITHIN THE ACCEPTABLE RANGE. THE FIELD SERVICE REPRESENTATIVE CHECKED THE INSTRUMENT AND PERFORMED PREVENTIVE MAINTENANCE. HE REPLACED THE PIPETTE NOZZLE AND THE BEAD MIXER PADDLE. HE INSTALLED A UPS SYSTEM TO AVOID ELECTRIC INTERFERENCE.