PRODIGY AUTOCODE
Report
- Report Number
- 3005862821-2017-00033
- Event Type
- Injury
- Date Received
- April 19, 2017
- Date of Event
- March 16, 2017
- Report Date
- March 16, 2017
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
BECAUSE DEVICE WAS NOT RETURNED TO OK BIOTECH, THEREFORE, WE WERE UNABLE TO PERFORM FURTHER TESTING ON THE SUSPECTED DEVICE. OK BIOTECH REVIEWED THE MANUFACTURING RECORD OF THIS SUSPECTED DEVICE (METER SERIAL NUMBER # (B)(4)), AND THE METER WAS QUALIFIED AND RELEASED BY THE QUALITY CONTROL DEPARTMENT AND SHIPPED TO (B)(4) ON 10/27/2015. WE ARE UNABLE TO CONFIRM THE COMPLAINT BECAUSE DEVICE WAS NOT RETURNED AND NO FURTHER INFORMATION FROM CUSTOMER HAS BEEN RECEIVED, THIS MATTER HAS TO BE CLOSED OUT WITH UNDETERMINED ROOT CAUSE.
SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. WE TESTED THE STANDBY CURRENT OF RETURNED METER, THE RESULT WAS 1.2UA. THE CRITERIA IS <55UA. PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. WE TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND IN HOUSE STRIPS (STRIP LOT NUMBER: D160526-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 59/54 MG/DL, FOR LEVEL HIGH WERE 259/260 MG/DL. THE REQUEST CONTROL SOLUTION RANGES ARE: LEVEL LOW 30~80 MG/DL; LEVEL HIGH 190~290 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.
IT WAS REPORTED THAT MEDICAL ATTENTION WAS SOUGHT ON (B)(6) 2017 AROUND 4 AM AFTER THE END USER RECEIVED INCONSISTENT RESULTS FROM THEIR PRODIGY DIABETES GLUCOSE METER. THE END USER WAS LETHARGIC, INCOHERENT AND UNRESPONSIVE. HIS BLOOD GLUCOSE READING AT THE TIME OF THE MEDICAL EVENT WAS 130 MG/DL. THE PARAMEDICS WERE CALLED AND PERFORMED A BLOOD GLUCOSE TEST WITH THEIR METER AND RECEIVED A RESULT OF 39 MG/DL. THE PARAMEDICS ADMINISTERED AN IV WITH FLUIDS ALONG WITH A GLUCOSE PASTE TO ASSIST IN STABILIZING HIS BLOOD GLUCOSE. NO FURTHER DETAILS WERE PROVIDED IN RELATION TO THIS MEDICAL EVENT.
THIS IS A SUPPLEMENTAL REPORT TO INITIAL REPORT 3005862821-2017-00033 TO SUBMIT INVESTIGATION RESULTS FROM THE MANUFACTURER FOR THE SUSPECT DEVICES. DEVICES WERE RETURNED FROM (B)(4) ON 05/03//2017 AND AN INVESTIGATION RESULTS OF THE SUSPECT DEVICES WERE COMPLETED BY OK BIOTECH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 287064 | PRODIGY AUTOCODE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51850 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention | LANTUS 54 UNITS AT NIGHT| NOVOLOG 25 UNITS |