COBAS 8000 C 702 MODULE
Report
- Report Number
- 1823260-2017-00820
- Event Type
- Malfunction
- Date Received
- April 19, 2017
- Date of Event
- March 31, 2017
- Report Date
- May 17, 2017
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K100853
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS EVENT OCCURRED IN (B)(6). UNIQUE IDENTIFIER (UDI)#: (B)(4).
INVESTIGATIONS HAVE CONFIRMED THAT THE OBSERVED DETERGENT CRYSTALLIZATION ON THE CUVETTE SEGMENTS CAN INFLUENCE THE BUN ASSAY, CAUSING LOWER VALUES TO BE GENERATED. THE CRYSTALS MAY MOVE FROM THE EDGE OF THE CUVETTE SEGMENT INTO THE FIRST OR LAST CUVETTE OF THAT SEGMENT. AN IMPACT MAY BE DETECTABLE FOR OTHER TESTS, BUT BUN IS HIGHLY SUSCEPTIBLE TO SUCH CRYSTALLIZATION. IT WAS RECOMMENDED TO THE CUSTOMER TO REMOVE THE CUVETTES COMPLETELY, CLEAN THE WATER BATH THOROUGHLY, INSTALL NEW CUVETTES, AND VERIFY PROBE ADJUSTMENTS.
THE CUSTOMER STATED THAT THEY HAVE BEEN HAVING ISSUES WITH INFREQUENT QUESTIONABLE LOW RESULTS FOR UREAL UREA/BUN (BUN) ON TWO COBAS 8000 C 702 MODULES (C702). AS PART OF TROUBLESHOOTING FOR THESE ISSUES IN THE PAST, IT WAS DETERMINED THAT A POSSIBLE CAUSE WAS LABORATORY HUMIDITY. IT WAS RECOMMENDED TO THE CUSTOMER AT THAT TIME TO USE A HUMIDIFIER IN THE LABORATORY. THIS MEDWATCH WILL APPLY TO THE SECOND AFFECTED C702 ANALYZER. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FIRST AFFECTED C702 ANALYZER. THE CUSTOMER STATES THAT THE ISSUE HAS NOW RETURNED STARTING ON (B)(6) 2017. THE CUSTOMER PROVIDED DATA FOR FOUR PATIENT SAMPLE THAT HAD QUESTIONABLE BUN RESULTS. OF THESE FOUR SAMPLES, ONE HAD AN ERRONEOUS BUN RESULT. IT WAS ASKED, BUT IT IS NOT KNOWN IF THE ERRONEOUS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED AS 5.6 MMOL/L AND WHEN REPEATED, IT RESULTED AS 23.5 MMOL/L. NO ADVERSE EVENTS WERE ALLEGED TO HAVE OCCURRED WITH THE PATIENT. THE BUN REAGENT LOT NUMBER WAS 212533. THE REAGENT EXPIRATION DATE WAS ASKED FOR, BUT NOT PROVIDED. THE FIELD SERVICE ENGINEER CHECKED THE PRE-ANALYTIC SAMPLE HANDLING CONDITIONS OF THE LABORATORY AND EVERYTHING WAS OK. THE SYSTEM WAS CHECKED AND IT WAS DETERMINED THAT THE CUVETTES WERE IN USE FOR 1.5 MONTHS. CRYSTALLIZATION WAS OBSERVED AT THE INTERFACE OF CUVETTE SEGMENTS. THE CUVETTES AND SAMPLE PROBES WERE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 285707 | COBAS 8000 C 702 MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | C702 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |