FDA Adverse Event
Malfunction
Summary report: N
TRIAGE METER PRO
MDR report key: 6491450
·
Received April 13, 2017
Report
- Report Number
- 2027969-2017-00066
- Event Type
- Malfunction
- Date Received
- April 13, 2017
- Date of Event
- March 16, 2017
- Report Date
- May 25, 2017
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- KHO
- PMA / PMN Number
- K973547
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER NARRATIVE: THE CUSTOMER'S COMPLAINT WAS NOT REPLICATED WITH THE RETURNED METER. NO EVIDENCE OF CHARRING OR BURNING WAS OBSERVED ON THE POWER SUPPLY OR METER. THE METER POWERED ON SUCCESSFULLY AND THE QCD PASSED. UNABLE TO DETERMINE A ROOT CAUSE OF THE COMPLAINT.
Additional Manufacturer Narrative · 1
INVESTIGATION PENDING.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A TRIAGE METER PRO POWER CORD BURNT THROUGH AFTER A POWER OUTAGE AT THE FACILITY. THERE WAS NO PATIENT OR USER INJURY. THE TRIAGE METER PRO WAS DAMAGED BY THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270862 | TRIAGE METER PRO | FLUOROMETER, FOR CLINICAL USE | KHO | ALERE SAN DIEGO, INC. | 55071 | 72419 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | METER POWER CORD, PN 52059 |