FDA Adverse Event Malfunction Summary report: N

TRIAGE METER PRO

MDR report key: 6491450 · Received April 13, 2017

Report

Report Number
2027969-2017-00066
Event Type
Malfunction
Date Received
April 13, 2017
Date of Event
March 16, 2017
Report Date
May 25, 2017
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
KHO
PMA / PMN Number
K973547
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER NARRATIVE: THE CUSTOMER'S COMPLAINT WAS NOT REPLICATED WITH THE RETURNED METER. NO EVIDENCE OF CHARRING OR BURNING WAS OBSERVED ON THE POWER SUPPLY OR METER. THE METER POWERED ON SUCCESSFULLY AND THE QCD PASSED. UNABLE TO DETERMINE A ROOT CAUSE OF THE COMPLAINT.

Additional Manufacturer Narrative · 1

INVESTIGATION PENDING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A TRIAGE METER PRO POWER CORD BURNT THROUGH AFTER A POWER OUTAGE AT THE FACILITY. THERE WAS NO PATIENT OR USER INJURY. THE TRIAGE METER PRO WAS DAMAGED BY THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270862 TRIAGE METER PRO FLUOROMETER, FOR CLINICAL USE KHO ALERE SAN DIEGO, INC. 55071 72419

Patients

Seq Age Sex Outcome Treatment
1 METER POWER CORD, PN 52059