FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION TREON SYSTEM

MDR report key: 649116 · Received November 21, 2005

Report

Report Number
649116
Event Type
Malfunction
Date Received
November 21, 2005
Date of Event
November 21, 2005
Report Date
November 21, 2005
Manufacturer
MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES
Product Code
HAW
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

WHEN PLUGGING IN THE STEALTHSTATION TREON, THE ELECTRICAL CIRCUIT BREAKER IN THE BUILDING TRIPPED, CAUSING LOSS OF POWER TO THE ANESTHESIA MACHINE AND THE SURGICAL LIGHTS. POWER FOR THE ANESTHESIA MACHINE WAS RESTORED BY PLUGGING INTO A DIFFERENT OUTLET. ANESTHETIC GAS MONITORING IS NOT AVAILABLE FOR THE FIRST 5 MINUTES AFTER A POWER LOSS, WHILE THE MONITOR RE-CALIBRATES.THIS PROBLEM IS RELATED TO AN EARLIER MEDSUN REPORT FROM THE SAME FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEALTHSTATION TREON SYSTEM SURGICAL NAVIGATION SYSTEM HAW MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES TREON *

Patients

Seq Age Sex Outcome Treatment
1 *