FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TREON SYSTEM
MDR report key: 649116
·
Received November 21, 2005
Report
- Report Number
- 649116
- Event Type
- Malfunction
- Date Received
- November 21, 2005
- Date of Event
- November 21, 2005
- Report Date
- November 21, 2005
- Manufacturer
- MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES
- Product Code
- HAW
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
Narratives
Description of Event or Problem · 1
WHEN PLUGGING IN THE STEALTHSTATION TREON, THE ELECTRICAL CIRCUIT BREAKER IN THE BUILDING TRIPPED, CAUSING LOSS OF POWER TO THE ANESTHESIA MACHINE AND THE SURGICAL LIGHTS. POWER FOR THE ANESTHESIA MACHINE WAS RESTORED BY PLUGGING INTO A DIFFERENT OUTLET. ANESTHETIC GAS MONITORING IS NOT AVAILABLE FOR THE FIRST 5 MINUTES AFTER A POWER LOSS, WHILE THE MONITOR RE-CALIBRATES.THIS PROBLEM IS RELATED TO AN EARLIER MEDSUN REPORT FROM THE SAME FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEALTHSTATION TREON SYSTEM | SURGICAL NAVIGATION SYSTEM | HAW | MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES | TREON | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |