FDA Adverse Event Injury Summary report: N

MO.MA

MDR report key: 6489707 · Received April 13, 2017

Report

Report Number
3004066202-2017-00040
Event Type
Injury
Date Received
April 13, 2017
Report Date
March 16, 2017
Manufacturer
INVATEC SPA
Product Code
NTE
PMA / PMN Number
K092177
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

INTRODUCTION: CAROTID ARTERY STENTING (CAS) IS AN ALTERNATIVE TREATMENT TO SURGICAL ENDARTERECTOMY IN SIGNIFICANT CAROTID ARTERY DISEASE. PROXIMAL PROTECTION DEVICES (PPD), ARE INCREASINGLY BEING USED FOR ACHIEVE CEREBRAL PROTECTION PRIOR TO CROSSING THE CAROTID STENOSIS. THIS STUDY EVALUATED THE SAFETY AND PERIOPERATIVE AND LONG-TERM EFFECTIVENESS IN PATIENTS WITH SIGNIFICANT CAROTID ARTERY STENOSIS TREATED WITH CAROTID STENTING AND PROXIMAL PROTECTION DEVICE (MEDTRONIC, (B)(4)) IN OUR CENTRE. METHODS: THIS PROSPECTIVE REGISTRY INCLUDED 125 PATIENTS (39.2% SYMPTOMATIC) WHO UNDERGONE CAS WITH PPD. THE EFFICACY WAS ASSESSED BY MEANS OF CUMULATIVE INCIDENCE OF ANY MAJOR ADVERSE EVENT (MAE) A COMPOSITE OF STROKE, MYOCARDIAL INFARCTION AND DEATH WITHIN 30 DAYS AFTER THE INTERVENTION OR IPSILATERAL STROKE AFTER 30 DAYS AND UP TO 4 YEARS. RESULTS: TECHNICAL SUCCESS WAS ACHIEVED IN 98.4% OF PROCEDURES. DURING THE PERIPROCEDURAL PERIOD (30 DAYS), THE INCIDENCE OF MAE WAS 1.6% ONE STROKE AND ONE DEATH, NOT RELATED WITH THE PROCEDURE. NO MYOCARDIAL INFARCTION WAS REPORTED. NO LONG-TERM IPSILATERAL STROKE OR NEUROLOGICAL DEATH WAS REPORTED. PROXIMAL ENDOVASCULAR OCCLUSION WAS WELL TOLERATED BY ALL OF THE PATIENTS. MINOR COMPLICATIONS INCLUDED 5 CASES (4%) OF TRANSIENT DISTAL SPASM OF INTERNAL CAROTID ARTERY (ALL OF THEM REASSUMED THEIR BASAL DIAMETER SPONTANEOUSLY WITHOUT LEADING TO ANY COMPLICATION) AND 1 CASE (0.8%) OF EXTERNAL CAROTID ARTERY THROMBOSIS. THERE WAS NO SIGNIFICANT DIFFERENCE IN THE RATES OF CUMULATIVE INCIDENCE OF MAE AT 30 DAYS BETWEEN SYMPTOMATIC AND ASYMPTOMATIC PATIENTS (0% VERSUS 2.6%; P¼0.52), NEITHER IN THE COMPONENTS, (STROKE 0% VERSUS1.3%; P¼0.6); AND DEATH (1.3% VERSUS 0%; P¼0.6). CONCLUSION: CAROTID ARTERY STENTING (CAS) WITH PROXIMAL PROTECTION DEVICE SEEMED EFFECTIVE AND SAFE IN OUR EXPERIENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272426 MO.MA CATHETER, CAROTID, TEMPORARY, FOR EMBOLIZATION CAPTURE NTE INVATEC SPA UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention