FDA Adverse Event Malfunction Summary report: N

THERMOPHORE

MDR report key: 64880 · Received January 23, 1997

Report

Report Number
1811605-1997-90020
Event Type
Malfunction
Date Received
January 23, 1997
Report Date
January 10, 1997
Manufacturer
BATTLE CREEK EQUIPMENT CO.
Product Code
IRT
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

THE SWITCH EMITTED SMOKE. CO'S INSPECTION REVEALED CARBON DEPOSITS WERE PRESENT IN THE SWITCH. NO DEATHS OR INJURIES WERE REPORTED. FROM APPROX. 1967 TO JUNE 21,1990, CO USED A LISTED SWITCH MANUFACTURED BY A REPUTABLE FIRM. THE SWITCH WAS A MOMENTARY TYPE SWITCH WITH A SPRING LOADED LEVER WHICH REQUIRED BEING HELD IN THE "ON" POSITION TO MAKE CONTACT. SOMETIMES A USER WOULD NOT FULLY ENGAGE THE LEVER, ALLOWING THE CONTACTS TO REMAIN SLIGHTLY OPEN WHICH COULD INDUCE ARCING. WHEN CONSISTENTLY HELD IN THIS WAY, CARBON BUILD-UP FROM THE ARCING COULD REACH OPPOSITE POLES CAUSING THE SWITCH TO EMIT A SMALL AMOUNT OF SMOKE. CO'S FIRST SOLUTION WAS TO CHANGE THE CONTACT MATERIAL AND THE BODY OF THE SWITCH. THIS REDUCED THE ARCING OF THE CONTACTS AND THE CARBON TRACKING OF THE SWITCH BODY. ALTHOUGH THE SITUATION WAS IMPROVED, THE OCCASIONAL RELEASE OF A PUFF OF SMOKE WAS POSSIBLE. ON JUNE 21, 1990, CO BEGAN USING A SNAP ACTION SWITCH WHICH DOES NOT ALLOW THE CONTACTS TO BE HELD IN A SLIGHTLY OPEN POSITION. THIS REPORT IS BEING MADE TO BE COMPLIANT WITH REGULATORY LETTER 90-DT-12. REMEDIAL ACTION HAS BEN ACCOMPLISHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE HEATING PAD IRT BATTLE CREEK EQUIPMENT CO. 055 NA

Patients

Seq Age Sex Outcome Treatment
1 *