FDA Adverse Event Malfunction Summary report: N

ETHERNET CAT5 TO FIBER CONV.

MDR report key: 6484908 · Received April 12, 2017

Report

Report Number
3005985723-2017-00174
Event Type
Malfunction
Date Received
April 12, 2017
Date of Event
March 7, 2017
Report Date
May 8, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K141989
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO CORRECT THE CATALOG #. REPORTED EVENT: ETHERNET CAT5 TO FIBER CONV ROBOT CONNECTION ERROR. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE ETHERNET CAT5 TO FIBER CONV EVENT WAS CONFIRMED. DEVICE HISTORY REVIEW: A REVIEW OF THE DHR ASSOCIATED WITH ROB343 FOUND QUALITY INSPECTION PROCEDURES SUCCESSFULLY PASSED WITH NO NOTES OR COMMENTS. COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING THE ETHERNET CAT5 TO FIBER CONV ROBOT CONNECTION ERROR FAILURE OF P/N: 200933. THERE HAVE BEEN NO OTHER SIMILAR EVENTS FOR THE REFERENCED SERIAL NUMBER. CONCLUSIONS: THE ETHERNET CAT5 TO FIBER CONV REPORTED EVENT WAS CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: AS THE EVENT DID NOT INVOLVE A MANUFACTURING RELATED PRODUCT PROBLEM INDICATING A NON-CONFORMITY, ADVERSE TREND, OR UNANTICIPATED HAZARD, NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

THE SURGEON WAS COMPLETING A TOTAL HIP ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM. THE MPS REPORTED A ROBOT CONNECTION ERROR. THE FIRST CASE WAS STARTED USING THE ROBOT BUT COMPLETED MANUALLY AND ALL OTHER CASES WERE COMPLETED USING THE BYPASS.

Description of Event or Problem · 1

THE SURGEON WAS COMPLETING A TOTAL HIP ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM. THE MPS REPORTED A ROBOT CONNECTION ERROR. THE FIRST CASE WAS STARTED USING THE ROBOT BUT COMPLETED MANUALLY AND ALL OTHER CASES WERE COMPLETED USING THE BYPASS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266340 ETHERNET CAT5 TO FIBER CONV. STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. ROB343

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization