FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6481590 · Received April 11, 2017

Report

Report Number
1000113657-2017-00645
Event Type
Malfunction
Date Received
April 11, 2017
Date of Event
March 16, 2017
Report Date
April 19, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00041220001286
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 243 AND 289 MG/DL. THE CUSTOMER WAS ALSO CONCERNED THAT THE RESULTS WERE ERRATIC. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 70 - 129 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/29/2018 AND OPEN VIAL DATE IS (B)(6) 2017. CUSTOMER STATED SHE TESTS FOUR TIMES A DAY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (TIME NOT SET): (B)(6). MEMORY CONCERNS: CUSTOMER IS CONCERN WITH THE RESULTS OF (B)(6) 2017. CUSTOMER STATES THAT HER SUGAR WAS DOING GOOD. CUSTOMER STATES THAT SHE RUN A BLOOD TEST AND GOT 289/243/278 MG/DL .CUSTOMER HAVE ONLY BEEN USING THIS METER SINCE (B)(6) 2017.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 243 AND 289 MG/DL. THE CUSTOMER WAS ALSO CONCERNED THAT THE RESULTS WERE ERRATIC. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 70 - 129 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/29/2018 AND OPEN VIAL DATE IS (B)(6) 2017. CUSTOMER STATED SHE TESTS FOUR TIMES A DAY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (TIME NOT SET): (B)(6). CUSTOMER STATES THAT HER SUGAR WAS DOING GOOD. CUSTOMER STATES THAT SHE RUN A BLOOD TEST AND GOT 289/243/278 MG/DL .CUSTOMER HAVE ONLY BEEN USING THIS METER SINCE (B)(6) 2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261892 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1987 00041220001286

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY