FDA Adverse Event Other Summary report: N

BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 648115 · Received November 30, 2005

Report

Report Number
2243072-2005-00471
Event Type
Other
Date Received
November 30, 2005
Date of Event
November 2, 2005
Report Date
November 30, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT INACCURATE READINGS WITH THE PT'S METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG READING (207) AS REFERENCE POINT VS. BD READINGS (331, 151, 139) YIELDED SOME DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064180

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other