FDA Adverse Event Malfunction Summary report: N

AXIOM ARTIS DBC

MDR report key: 6480263 · Received April 11, 2017

Report

Report Number
2240869-2017-01905
Event Type
Malfunction
Date Received
April 11, 2017
Date of Event
March 23, 2017
Report Date
March 23, 2017
Manufacturer
SIEMENS HEALTHINEERS, BUS. AREA ADVANCED THERAPY
Product Code
OWB
PMA / PMN Number
K052202
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SIEMENS HAS COMPLETED AN INVESTIGATION OF THE REPORTED EVENT. THE PROBLEM OCCURRED DUE TO AN ERROR WITHIN THE BSR (IMAGE SYSTEM X-RAY). LOG FILE ASSESSMENT OF PLANE A DOES NOT SHOW A FAILURE. THE IMAGE QUALITY OF PLANE B WAS INVESTIGATED. DURING THE PERIOD OF INTEREST, THE IMAGE QUALITY FOR THAT PERIOD WAS SATISFACTORY. A PROBLEM WITH THE BSR (IMAGE SYSTEM X-RAY) WAS FOUND ON PLANE B. ONLY LIMITED IMAGE QUALITY WAS AVAILABLE FOR APPROXIMATELY 9 MINUTES. AFTER AUTO-RECOVERY, FULL IMAGE QUALITY WAS AGAIN AVAILABLE. THERE WAS NO LIMITATION AND PLANE A WAS FULLY OPERATIONAL. THE SYSTEM PERFORMED A SELF-RECOVERY AND IS OPERATING AS SPECIFIED. FAILURE OF THE BSR IS BEING MONITORED VIA THE CAPA PROCESS. IN CASE OF AN INCREASE, FURTHER ACTIONS ARE TO BE CONSIDERED.

Additional Manufacturer Narrative · 1

SIEMENS IS CONDUCTING A THOROUGH INVESTIGATION OF THE REPORTED EVENTS. AS THIS EVENT IS UNDER INVESTIGATION, A ROOT CAUSE HAS NOT YET BEEN DETERMINED. A SUPPLEMENT REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE AXIOM ARTIS DBC SYSTEM. DURING A HEART VALVE IMPLANTATION PROCEDURE, THE FLUORO A PLANE IMAGE BECAME TOO BRIGHT. A SYSTEM RESTART WAS ATTEMPTED AT WHICH POINT AN ERROR MESSAGE WAS RECEIVED. THE PATIENT WAS SAFELY REMOVED FROM THE SYSTEM AND TRANSFERRED TO AN ALTERNATIVE SYSTEM WHERE THE PROCEDURE WAS COMPLETED. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261132 AXIOM ARTIS DBC INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB SIEMENS HEALTHINEERS, BUS. AREA ADVANCED THERAPY 07728392

Patients

Seq Age Sex Outcome Treatment
1