X100L BD ULTRASAFE PASSIVE¿ X-SERIES NEEDLE GUARD SYRINGE
Report
- Report Number
- 3009081593-2017-00005
- Event Type
- Malfunction
- Date Received
- April 7, 2017
- Date of Event
- February 27, 2017
- Report Date
- April 7, 2017
- Manufacturer
- BECTON DICKINSON HUNGARY KFT (BD)
- Product Code
- MEG
- PMA / PMN Number
- K011369
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
RESULTS: A SAMPLE IS NOT AVAILABLE FOR EVALUATION. AS PART OF A NO SAMPLE INVESTIGATION, 36 RETENTION SAMPLES WERE EVALUATED FOR SAFETY MECHANISM FAILURE; 10 UNITS WERE TESTED BY ACTIVATION CYCLE AND 26 UNITS WERE TESTED BY MANUAL TRIGGERING. ALL SAMPLES PASSED TESTING, REACHED LOCK-OUT POSITION AFTER ACTIVATION, AND DID NOT SHOW ANY SAFETY DEVICE ISSUES. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES OR NON-CONFORMANCES DURING THE MANUFACTURE OF THE REPORTED LOT # 5243391. A MANUFACTURING PROCESS REVIEW SHOWED NO ISSUES WITH SPRING FORMATION, DEVICE ASSEMBLY, PACKAGING, RAW MATERIAL INCOMING INSPECTION, IN PROCESS INSPECTION (100% AUTOMATED PRIMARY FUNCTION TEST, CONTROLLED TRIGGER TEST, ACTIVATION CYCLE TEST), OR FINAL INSPECTION. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. NO ISSUES WERE OBSERVED WITH RETENTION SAMPLE TESTING OR FOUND WITHIN THE DEVICE HISTORY RECORD AND MANUFACTURING PROCESS REVIEWS. (B)(4).
IT WAS REPORTED THAT AFTER A PATIENT HAD INJECTED HERSELF USING A X100L BD ULTRASAFE PASSIVE¿ X-SERIES NEEDLE GUARD SYRINGE, THE SAFETY MECHANISM FAILED TO FUNCTION PROPERLY. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253239 | X100L BD ULTRASAFE PASSIVE¿ X-SERIES NEEDLE GUARD SYRINGE | PISTON SYRINGE | MEG | BECTON DICKINSON HUNGARY KFT (BD) | 5243391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |