NEWPORT MEDICAL INSTRUMENTS, INC
Report
- Report Number
- 2023050-2017-05135
- Event Type
- Malfunction
- Date Received
- April 5, 2017
- Report Date
- June 13, 2017
- Manufacturer
- NEWPORT MEDICAL INSTRUMENTS, INC.
- Product Code
- CBK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT ANALYSIS #225066432: UNABLE TO DUPLICATED FAILURE. THE OXYGEN SENSOR WAS INSTALLED INTO THE TEST UNIT TO THE VERIFY SYMPTOM AND PASSED TESTING. THE SETTINGS WERE ADJUSTED TO 30%, 50%, 80% AND 100%. THE TOLERANCE FOR THE OXYGEN SENSOR SHOULD NOT EXCEED +/- 3%. THE FOLLOWING MEASUREMENTS WERE OBTAINED FROM THE SERVOMEX 30.1%, 50.3%, 80.3%, AND 99.5%.
IT WAS REPORTED THAT THE VENTILATOR'S OXYGEN READINGS WERE HIGHER THAN SET VALUE. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER DECLINED TO PROVIDE THE TYPE OF THE DEVICE.
IT WAS REPORTED THAT THE VENTILATOR'S OXYGEN READINGS WERE HIGHER THAN SET VALUE. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER DECLINED TO PROVIDE THE TYPE OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242290 | NEWPORT MEDICAL INSTRUMENTS, INC | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | NEWPORT MEDICAL INSTRUMENTS, INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |