METRIX AIR
Report
- Report Number
- 1000113657-2017-00618
- Event Type
- Malfunction
- Date Received
- April 5, 2017
- Date of Event
- March 13, 2017
- Report Date
- April 5, 2017
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292007010
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). (B)(4).
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 151, 172 AND 174 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 120 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 11/30/2017 AND OPEN VIAL DATE IS (B)(6) 2017. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE NOT SET; MEMORY CONCERNS: CUSTOMER CONCERNED WITH RESULTS DATE (B)(6) 2017 AT 12:56 PM): (B)(6). MEMORY CONCERNS: CUSTOMER CONCERNED WITH RESULTS DATE (B)(6) 2017 AT 12:56 PM. THE CUSTOMER CALLED WITH CONCERNS REGARDING THEIR METER READING HIGH. THEIR NORMAL FASTING GLUCOSE RANGE IS 100MG/DL TO 120MG/DL. THE CUSTOMER DENIES THE NEED FOR MEDICAL ATTENTION AT THE TIME OF CALL. THE CUSTOMER DENIES HAVING ANY SIGNS OR SYMPTOMS OF HYPOGLYCEMIA OR HYPERGLYCEMIA AT THE TIME OF CALL. THE METER'S DATE IS NOT SET UP CORRECTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242761 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MT1821 | 00021292007010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |