FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME
Report
- Report Number
- 1037905-2017-00143
- Event Type
- Malfunction
- Date Received
- April 3, 2017
- Report Date
- April 3, 2017
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- KNS
- UDI-DI
- 00827002315334
- PMA / PMN Number
- K052051
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NURSE
Narratives
INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE CUTTING WIRE SECURING COMPONENT LOCATED NEAR THE DISTAL END OF THE SPHINCTEROTOME HAS DISCONNECTED FROM THE CATHETER. THE CUTTING WIRE IS INTACT AND REMAINS SECURELY ATTACHED TO THE SPHINCTEROTOME AT THE PROXIMAL END. HOWEVER, DUE TO THE CATHETER AND SECURING COMPONENT DISCONNECTION, THE DISTAL END OF THE CUTTING WIRE IS NO LONGER CONNECTED TO THE SPHINCTEROTOME CATHETER AT THE DISTAL END. THE SECURING COMPONENT HAS A LONGER SECTION MEASURING 3 MM AND A SHORTER SECTION MEASURING 2 MM. NO SECTION OF THE SPHINCTEROTOME DEVICE IS MISSING. A VISUAL EXAMINATION OF THE CATHETER SHOWED NO KINKS/BENDS. THE WIRE GUIDE PRELOADED INTO THIS DEVICE WAS NOT INCLUDED IN THE RETURN OF THE DEVICE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER: "DO NOT OVER FLEX OR BOW TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE OR CAUSE CUTTING WIRE TO BREAK." THIS TYPE OF DAMAGE CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRE-CURVED AND IS PROVIDED WITH A PRE-CURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: ¿NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF SPHINCTEROTOME IN AN ATTEMPT TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE.¿ IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADDITIONAL PRESSURE IS APPLIED, THIS COULD HAVE CONTRIBUTED TO SEPARATION OF THE CATHETER AND CUTTING WIRE SECURING COMPONENT. THE INSTRUCTIONS FOR USE CAUTION THE USER: "ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING SPHINCTEROTOME." THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER: ¿UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME." THE USER IS THEN INSTRUCTED: "CAREFULLY REMOVE PRECURVED STYLET WIRE FROM CANNULATING TIP.¿ THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." THE INSTRUCTIONS FOR USE ADVISE THE USER: ¿BEFORE USING THIS DEVICE, FOLLOW RECOMMENDATIONS PROVIDED BY ELECTROSURGICAL UNIT MANUFACTURER TO ENSURE PATIENT SAFETY THROUGH PROPER PLACEMENT AND UTILIZATION OF PATIENT RETURN ELECTRODE. ENSURE A PROPER PATH FROM PATIENT RETURN ELECTRODE TO ELECTROSURGICAL UNIT IS MAINTAINED THROUGHOUT PROCEDURE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. THE FUNCTIONAL TEST INCLUDES BOWING THE SPHINCTEROTOME TO ENSURE THE DISTAL END RESPONDS TO HANDLE MANIPULATION. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: " IN THE SAME PROCEDURE, THE CUTTING WIRE DETACHED [CUTTING WIRE SECURING COMPONENT SEPARATED FROM CATHETER] FROM THE PROXIMAL END OF THE CATHETER AND WAS UNUSABLE. A THIRD DEVICE ALLOWED THEM TO COMPLETE THE PROCEDURE." [SEE RELATED MDR 1037905-2017-00142].
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236784 | FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME | KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | COOK ENDOSCOPY | 00827002315334 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS ENDOSCOPE (UNKNOWN MODEL) |