VERSAFITCUP FLAT PE LINER Ø 28 / G
Report
- Report Number
- 3005180920-2017-00150
- Event Type
- Injury
- Date Received
- March 31, 2017
- Date of Event
- March 3, 2017
- Report Date
- March 31, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- PMA / PMN Number
- K103352
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEWS PERFORMED ON 27 MARCH 2017. LOT 05-0701: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25 AUGUST 2005. EXPIRATION DATE: 2010-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 5 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. QUADRA-H CEMENTLESS, HA COATED STEM SIZE 8 STD, CODE 01.12.028, LOT. 070663 (K082792) LOT 070663: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01 JUNE 2007. EXPIRATION DATE: 2012-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD AND THIS IS THE SECOND SIMILAR EVENT REPORTED ON THE SUBJECT LOT. ON 30 MARCH 2017 THE MEDICAL AFFAIRS DIRECTOR PERFORMED A CLINICAL EVALUATION BASED ON THE IMAGE PROVIDED BY THE REPORTER AND COMMENTED AS FOLLOWS: REVISION OF STANDARD (NON-CROSSLINKED) POLYETHYLENE LINER 9 YEARS AFTER PRIMARY CEMENTLESS THA IN A (B)(6) MAN. RADIOGRAPHIC IMAGE SHOWS THE PRESENCE OF REGIONS OF OSTEOLYSIS IN THE PROXIMAL PART OF THE FEMUR, WHICH ARE LIKELY DUE TO ACCUMULATION OF PE DEBRIS. WEAR OF POLYETHYLENE IS A POSSIBLE, LITERATURE DESCRIBED, EVENT THAT CAN OCCUR IN THE LONG-TERM AFTER PRIMARY THA. TRIBOLOGICAL PAIRINGS THAT OFFER GOOD VERY-LONG-TERM RESULTS ARE AVAILABLE AND MAY BE A PREFERRED CHOICE FOR PATIENTS VERY YOUNG AND ACTIVE (THIS PATIENT WAS (B)(6) AT THE AGE OF PRIMARY OPERATION).
REVISION OF THE INLAY AND THE HEAD PLANNED DUE TO OSTEOLYSES IN PROXIMAL ZONE AND POSSIBLE WEAR OF THE LINER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 231644 | VERSAFITCUP FLAT PE LINER Ø 28 / G | FLAT PE LINER | LZO | MEDACTA INTERNATIONAL SA | 05-0701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention |