FDA Adverse Event Injury Summary report: N

PIPELINE EMBOLIZATION DEVICE

MDR report key: 6448706 · Received March 30, 2017

Report

Report Number
2029214-2017-00309
Event Type
Injury
Date Received
March 30, 2017
Date of Event
January 6, 2017
Report Date
March 3, 2017
Manufacturer
COVIDIEN (IRVINE)
Product Code
OUT
PMA / PMN Number
P100018.S004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CITATION: MATTHEW B. POTTS, MD, MAKSIM SHAPIRO, MD, DANIEL W. ZUMOFEN, MD, EYTAN RAZ, MD, EREZ NOSSEK, MD, KEITH G. DESOUSA, MD, TIBOR BECSKE, MD, HOWARD A. RIINA, MD, AND PETER K. NELSON, MD. PARENT VESSEL OCCLUSION AFTER PIPELINE EMBOLIZATION OF CEREBRAL ANEURYSMS OF THE ANTERIOR CIRCULATION. J NEUROSURG JANUARY 6, 2017. THE DEVICE WAS NOT RETURNED AS IT WAS IMPLANTED IN THE PATIENT; THEREFORE, PRODUCT ANALYSIS OF THE PIPELINE DEVICES WAS NOT ABLE TO BE PERFORMED. THE MODEL AND LOT NUMBERS OF THE PIPELINE EMBOLIZATION DEVICES IMPLANTED WERE NOT REPORTED IN THE ARTICLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE MAIN AUTHOR OF THIS ARTICLE, HOWEVER NO FURTHER INFORMATION HAS BEEN PROVIDED. SHOULD THE REQUESTED INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE AUTHORS PERFORMED A RETROSPECTIVE REVIEW OF ALL ANTERIOR CIRCULATION ANEURYSMS CONSECUTIVELY TREATED AT A SINGLE INSTITUTION WITH THE PED THROUGH 2014, IDENTIFYING THOSE WITH PVO ON FOLLOW-UP IMAGING. THE CAUSES OF PVO IN THIS SERIES WERE NOT CONFIDENTLY IDENTIFIED BY THE AUTHORS. THE AUTHORS NOTE THAT THE CASES OF DELAYED PVO IN THE LITERATURE HAVE BEEN REPORTED TO OCCUR SOON AFTER DISCONTINUATION OF CLOPIDOGREL, EITHER AS PRESCRIBED BY THE TREATING PHYSICIANS BASED ON MEDICAL NECESSITY OR FOLLOW-UP IMAGING 6,9,13 OR WHEN SELF-DISCONTINUED BY PATIENT. BASED ON THE REPORTED INFORMATION, THERE DID NOT APPEAR TO HAVE BEEN ANY DEFECT OF THE DEVICE DURING USE. AFTER REVIEWING THE IFU, AND REVIEW OF LITERATURE TO INVESTIGATE THE COMPLAINT, THE MOST LIKELY CAUSE FOR THE COMPLAINT IS RELATED TO THE PATIENT CONDITION. REFERENCE MDRS 2029214-2017-00308, 2029214-2017-00309, 2029214-2017-00310 FOR THIS SUBJECT. OTHER MDRS FROM THIS LITERATURE REVIEW: 2029214-2017-00295, 2029214-2017-00296, 2029214-2017-00297, 2029214-2017-00298, 2029214-2017 -00299, 2029214-2017-00300, 2029214-2017-00301, 2029214-2017-00302, 2029214-2017-00303, 2029214-2017-00304, 2029214-2017-00305, 2029214-2017-00306, 2029214-2017-00307, 2029214-2017-00311, 2029214-2017-00312, 2029214-2017-00313, 2029214-2017-00314, 2029214-2017-00315, 2029214-2017-00316, 2029214-2017-00317, 2029214-2017-00318, 2029214-2017 -00319, 2029214-2017-00320, 2029214-2017-00321, 2029214-2017-00322 2029214-2017-00323, 2029214-2017-00324, 2029214-2017-00325, 2029214-2017-00326, 2029214-2017-00327, 2029214-2017-00328, 2029214-2017 -00329.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THROUGH LITERATURE REVIEW THAT THIS PATIENT EXPERIENCED ASYMPTOMATIC PARENT VESSEL OCCLUSION (PVO) AFTER PIPELINE EMBOLIZATION PROCEDURE. THE PVO WAS DISCOVERED ON ROUTINE FOLLOW-UP IMAGING AT 6 MONTHS. THERE WAS NO NEW PERMANENT NEUROLOGICAL DEFICIT AND THE GOAL OF COMPLETE ANEURYSM OCCLUSION WAS ACHIEVED. THERE WAS COLLATERAL SUPPLY TO THE OCCLUDED ICA TERRITORY THROUGH THE CIRCLE OF WILLIS¿VIA THE ANTERIOR COMMUNICATING ARTERY AND THE IPSILATERAL POSTERIOR COMMUNICATING ARTERY. THIS (B)(6) PATIENT PRESENTED WITH PROPTOSIS AND WAS FOUND TO HAVE A CAVERNOUS ANEURYSM IN THE LEFT INTERNAL CAROTID ARTERY (ICA) MEASURING 22MM WITH A NECK WIDTH OF 10.2MM. THIS PATIENT WAS TREATED WITH 3 PIPELINE. THE PATIENT REPORTED A MILD HEADACHE WITHIN DAYS AFTER THE EMBOLIZATION PROCEDURE THAT WAS INITIALLY TREATED WITH STEROIDS. THIS HEADACHE PERSISTED FOR SEVERAL MONTHS BEFORE THE DELAYED DISCOVERY OF THE PVO. THIS PATIENT ALSO UNDERWENT 5 YEAR ANGIOGRAPHIC FOLLOW-UP THAT DEMONSTRATED PERSISTENT PARENT VESSEL AND ANEURYSM OCCLUSIONS. THIS PATIENT RECEIVED CLOPIDOGREL, 75 MG DAILY, FOR 7 DAYS PRIOR TO EMBOLIZATION AND ASPIRIN, 325 MG DAILY, FOR 2 DAYS PRIOR TO EMBOLIZATION WITHOUT PRE- OR POST-PROCEDURAL PLATELET FUNCTION TESTING. THIS PROCEDURE WAS PERFORMED PRIOR TO P2Y12 PLATELET FUNCTIONAL TESTING BECAME AVAILABLE. UPON DISCHARGE THE PATIENT WAS PRESCRIBED A MINIMUM OF 6 MONTHS OF DUAL ANTIPLATELET THERAPY (DAPT) WITH ASPIRIN AND CLOPIDOGREL. THE DECISION TO STOP OR EXTEND THE COVERAGE WAS SUBSEQUENTLY BASED ON THE RESULTS OF FOLLOW UP ANGIO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230584 PIPELINE EMBOLIZATION DEVICE INTRACRANIAL ANEURYSM FLOW DIVERTER OUT COVIDIEN (IRVINE) UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 61 YR Other