FDA Adverse Event Other Summary report: N

EMPOWERCT CEILING COLUMN

MDR report key: 644853 · Received November 3, 2005

Report

Report Number
2411512-2005-00007
Event Type
Other
Date Received
November 3, 2005
Date of Event
October 3, 2005
Report Date
October 3, 2005
Manufacturer
ALGONQUIN INDUSTRIES, INC
Product Code
JAK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE E-Z-EM FIELD SERVICE TECHNICIAN RECEIVED A CALL FROM THE USER FACILITY STATING THE COMPLETE CEILING MOUNT SYSTEM FELL. PER THE FACILITY ENGINEER, THE ANCHORS INSTALLED TO SECURE THECEILING COLUMN'S PLATE TO THE CEILING FAILED. FACILITY ENGINEER REQUESTED AN INSTALLATION MANUAL AND INSTALLED THE INJECTOR ON THEIR ORIGINAL FLOOR STAND. THE INJECTOR WAS TESTED AND IS WORKING PROPERLY. NO ONE WAS INJURED AS A RESULT OF THIS EVENT, ALTHOUGH THE TECHNICIAN INDICATED SHE STRAINED HER MUSCLES HOLDING THE SYSTEM UP SO IT WOULD NOT FALL ON THE PATIENT. NO SUBSEQUENT MEDICAL INTERVENTION WAS NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EMPOWERCT CEILING COLUMN CEILING MOUNT ACCESSORY JAK ALGONQUIN INDUSTRIES, INC NA NA

Patients

Seq Age Sex Outcome Treatment
1 52 YR Other