INTRA LUX HEAD 68 LDN
Report
- Report Number
- 3003637274-2017-00020
- Event Type
- Injury
- Date Received
- March 30, 2017
- Report Date
- March 30, 2017
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EGS
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE PRODUCT HAS BEEN SOLD IN 06/1995 TO THE CUSTOMER. SINCE THAT TIME, IT HAS NEVER BEEN RETURNED FOR A SERVICE CHECK WHICH IS REQUIRED ON A REGULAR BASE DEPENDING ON THE FREQUENCY OF USE. UNFORTUNATELY, CUSTOMER IS NOT WILLING TO SUPPLY A REPAIR HISTORY. ONLY THE HEAD HAS BEEN AVAILABLE FOR THE ANALYSIS, THE SWALLOWED BACK CAP HAS BEEN MISSING. HOWEVER THIS WOULD HAVE BEEN A VERY IMPORTANT PART TO GET A COMPLETE VIEW OF THE SITUATION. THE HEAD HAD SEVERAL NICKS ALL AROUND WHICH INDICATES THAT IT RECEIVED EXTERNAL FORCE, E.G. HITS. SUCH HITS COULD CAUSE THAT THE BACK CAP GETS UNLOCKED. IF IT GETS USED FURTHER THE VIBRATIONS COULD CAUSE THAT IT UNSCREWS FURTHER AND DROPS FINALLY OFF THE HEAD. IF THE FUNCTIONAL AND VISUAL TEST IS DONE PRIOR TO EACH TREATMENT AS REQUESTED BY THE IFU IT WOULD BE RECOGNIZABLE THAT THE BACK CAP IS LOOSE AND SHOULD BE RE-TIGHTENED AGAIN. INCIDENT GETS REPORTED AS SIMILAR PRODUCTS GET SOLD IN USA.
IT WAS DESCRIBED THAT DURING A STANDARD DENTAL TREATMENT THE BACK CAP OF THE INSTRUMENT CAME LOOSE AND FELL INTO PATIENTS MOUTH. PATIENT THEN SWALLOWED THE BACK CAP. AN X-RAY SHOWED THAT IT WAS IN THE STOMACH. NO FURTHER MEDICAL CARE NECESSARY. THE DENTAL OFFICE DIDN'T WANT TO SUPPLY FURTHER DATA, HENCE DATE OF EVENT AND PATIENT DATA ARE NOT KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230280 | INTRA LUX HEAD 68 LDN | DENTAL HEAD ATTACHMENT | EGS | KAVO DENTAL GMBH | 68 LDN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |