FDA Adverse Event Malfunction Summary report: N

MICROSCAN DRIED POS MIC PANEL TYPE 20A

MDR report key: 644480 · Received September 27, 2005

Report

Report Number
2919016-2005-00025
Event Type
Malfunction
Date Received
September 27, 2005
Date of Event
August 29, 2005
Report Date
August 29, 2005
Manufacturer
DADE BEHRING, INC.
Product Code
LTT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SINGLE ACCOUNT REPORTED TO DADE BEHRING THAT THEY OBSERVED A CLINICAL S. AUREUS ISOLATE OXACILLIN MIC DISCREPANCY. THE ACCOUNT OBTAINED AN OXACILLN-SUSCEPTIBLE RESULT ON THE MICROSCAN DRIED POS MIC PANEL TYPE 20A LOT#2006-06-10 AND AN OXACILLI-RESISTANT RESULT ON A SECONDARY METHOD (MRSA SCREEN PLATE). ISOLATE WAS REPORTED AS "RESISTANT" BY THE LABORATORY TO THE PHYSICIAN. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES TO THE PT AS A RESULT OF ADVERSE HEALTH CONSEQUENCES TO THE PT AS A RESULT OF THE FALSESUSCEPTIBLE RESULT BEING OBTAINED. ISOLATES WERE SENT BY THE ACCOUNT TO DADE BEHRING FOR INVESTGATION.