PIPELINE EMBOLIZATION DEVICE
Report
- Report Number
- 2029214-2017-00236
- Event Type
- Malfunction
- Date Received
- March 28, 2017
- Date of Event
- June 8, 2015
- Report Date
- March 3, 2017
- Manufacturer
- COVIDIEN (IRVINE)
- Product Code
- OUT
- PMA / PMN Number
- P100018.S004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
CITATION: LI-MEI LIN,1 GEOFFREY P COLBY,2 BOWEN JIANG ET. AL INTRA-DIC (DISTAL INTRACRANIAL CATHETER) DEPLOYMENT OF THE PIPELINE EMBOLIZATION DEVICE: A NOVEL RESCUE STRATEGY FOR FAILED DEVICE EXPANSION. J NEUROINTERVENT SURG 2015;0:1¿7. DOI:10.1136/NEURINTSURG-2015-011771 1 THE PIPELINE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION AS IT WAS IMPLANTED IN THE PATIENT; THEREFORE, NO DEFINITIVE CONCLUSION CAN BE DRAWN REGARDING THE CLINICAL OBSERVATION. ADDITIONAL INFORMATION WAS REQUESTED, HOWEVER, NO FURTHER INFORMATION WAS PROVIDED BY THE AUTHOR OF THE ARTICLE. MDRS RELATED TO THIS EVENT: 2029214-2017-00234 2029214-2017-00235 2029214-2017-00236 2029214-2017-00237 2029214-2017-00238 2029214-2017-00239 2029214-2017-00240 2029214-2017-00241 2029214-2017-00242 2029214-2017-00243 2029214-2017-00244.
MEDTRONIC RECEIVED INFORMATION DURING LITERATURE REVIEW THAT THE PIPELINE FAILED TO EXPAND DURING EMBOLIZATION TREATMENT OF SMALL PARA OPHTHALMIC ANEURYSM. IT WAS REPORTED THAT THE PIPELINE WAS OPENED AND WELL APPOSED AT ITS DISTAL END IN THE SUPRACLINOID INTERNAL CAROTID ARTERY. HOWEVER, THE LARGER DIAMETER PIPELINE HAD DIFFICULTY IN OPENING AROUND THE POSTERIOR GENU AND FAILED TO FULLY EXPAND. THE DISTAL INTRACRANIAL CATHETER TECHNIQUE WAS THEN USED TO COMPLETE THE DEPLOYMENT OF THE PROXIMAL END. THE FINAL DEPLOYED PIPELINE WAS WELL APPOSED AND SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222960 | PIPELINE EMBOLIZATION DEVICE | INTRACRANIAL ANEURYSM FLOW DIVERTER | OUT | COVIDIEN (IRVINE) | FA-77450-16 | NOT REPORTED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |