FDA Adverse Event Malfunction Summary report: N

MARATHON DR

MDR report key: 64380 · Received January 16, 1997

Report

Report Number
1640319-1997-00030
Event Type
Malfunction
Date Received
January 16, 1997
Date of Event
December 19, 1996
Report Date
January 16, 1997
Manufacturer
INTERMEDICS, INC.
Product Code
DXY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

USER PRESETS FOR SR AND DR MODELS ARE STORED IN THE SAME LOCATION IN THE RX5000 PROGRAMMER. WHEN RECALLED, THE LAST SAVED WILL BE DISPLAYED (UNLESS USER PRESETS HAVE BEEN PREVIOUSLY STORED FOR THE MODEL).

Description of Event or Problem · 1

THE RPTR WAS SETTING UP USER PRESET VALUES IN THE RX5000 PROGRAMMER FOR A CO DR AND SR PACEMAKER, WHEN HE FOUND THE FOLLOWING. HE SET UP PUP VALUES FOR SR USING A LIVE DEVICE. HE THEN WENT INTO THE MARATHON DR SCREENS VIA SF1 FROM THE MAINS SCREEN, AND SET UP VALUES FOR THE DR PUP. AGAIN USING A LIVE SR, HE INQUIRED THEM AND RESTORED PUP VALUES. THE SCREEN SHOWED DDDR PUP VALUES, WHICH ARE NOT PROGRAMMABLE IN SR. FURTHER EVENT DISCRIPTION IS ON PAGE 3 OF THIS FORM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MARATHON DR Implant IMPLANTABLE PACEMAKER DXY INTERMEDICS, INC. 294-09 NA

Patients

Seq Age Sex Outcome Treatment
1 *