FDA Adverse Event Malfunction Summary report: N

RETROGRADE DRL 9MM

MDR report key: 6431871 · Received March 24, 2017

Report

Report Number
1219602-2017-00336
Event Type
Malfunction
Date Received
March 24, 2017
Date of Event
January 25, 2017
Report Date
April 6, 2017
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GFG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

VISUAL ASSESSMENT OF THE DEVICE CONFIRMED THE REPORTED COMPLAINT OF BREAKAGE. THE DRILL HEAD HAS SPLIT IN TWO PIECES AND HAS COME FREE FROM THE DRILL SHAFT. THE BREAK IS A CLEAN SHEAR WITH A SLIGHT TWIST. THE ACTUATOR WIRE IS SPIRALED IN THE FORWARD DRILLING POSITION. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED WHICH CONFIRMED NO INCONSISTENCIES. A ROOT CAUSE INVESTIGATION HAS BEEN OPENED TO ADDRESS THIS FAILURE MODE. AFTER THE EVALUATION THE ROOT CAUSE FOR THE REPORTED ISSUE COULD NOT BE DETERMINED. THERE ARE NO INDICATIONS THAT WOULD SUGGEST THAT THE DEVICE DID NOT MEET PRODUCT SPECIFICATIONS UPON RELEASE INTO DISTRIBUTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ACL REPAIR, THE BLADE POPPED OFF AND FELL IN THE KNEE. NO REPORTED PATIENT INJURIES. NO OTHER COMPLICATIONS WERE NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215393 RETROGRADE DRL 9MM BIT, SURGICAL GFG SMITH & NEPHEW, INC.

Patients

Seq Age Sex Outcome Treatment
1 19 YR