FDA Adverse Event Injury Summary report: N

ECLIPSE TRUNION,49 MM TPS CTD

MDR report key: 6431858 · Received March 24, 2017

Report

Report Number
1220246-2017-00093
Event Type
Injury
Date Received
March 24, 2017
Date of Event
February 15, 2017
Report Date
March 27, 2017
Manufacturer
ARTHREX, INC.
Product Code
KWS
PMA / PMN Number
K083435
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THIS IS ONE OF FOUR SUBMISSIONS FOR THE SAME EVENT. THE OTHERS ARE (B)(4). THE DEVICE WAS RECEIVED AND AN EVALUATION WAS CONDUCTED. DEVICE HISTORY RECORD REVIEW REVEALED NOTHING RELEVANT TO THIS EVENT. THE EVALUATION OF THE RETURNED DEVICE REVEALED EXTENSIVE DAMAGES AND DISRUPTED SURFACE FINISH. THE DAMAGES OBSERVED ARE CONSISTENT WITH EXTRACTION MARKS. THE DEVICE'S CRITICAL MATING FEATURES ARE WITHIN SPECIFICATION. AT THIS TIME, IT CANNOT BE DETERMINED HOW THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE PATIENT'S EXPERIENCE. THIS IS THE FIRST COMPLAINT OF THIS TYPE FOR THIS PART/LOT COMBINATION. THE POTENTIAL CAUSE(S) OF THIS EVENT WILL BE COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT HAD ORIGINAL ECLIPSE SURGERY IN 2013. GLENOID AR-9105-03, (LOT 1230011) ((B)(4)), SCREW-AR-9301-02, (LOT 1127024) ((B)(4)), TRUNION-AR-9300-49CPC, (LOT 1128006) ((B)(4)), HUMERAL HEAD- AR-9349-18, (LOT 947018) ((B)(4)) WERE IMPLANTED. FOLLOW-UP SERIAL X RAYS SHOWED INCREASING LUCENT LINES AROUND THE PEGGED GLENOID. PATIENT HAD PAIN AND DECISION WAS MADE TO REVISE TO A REVERS PER SURGEON. REVISION TOOK PLACE ON (B)(6) 2017. THE HUMERAL COMPONENT WAS REMOVED WITHOUT ISSUE. SALES REP NOTED THAT THE ECLIPSE PROSTHESIS WAS EXTREMELY WELL FIXED AND THE CAGE SCREW HAD COMPLETE BONE THROUGH GROWTH. THE GLENOID WAS LOOSE AND WAS REMOVED BY HAND WITH PEGS INTACT. CEMENT WAS REMOVED FROM THE GLENOID. AUTOGRAFT WAS TAKEN FROM THE ILIAC CREST TO ADDRESS THE BONE DEFECT IN THE GLENOID. THE ALLOGRAFT WAS PREPARED FOR THE REVERS BASE PLATE IN SITU PRIOR TO HARVESTING THE GRAFT. THE BASEPLATE WAS THEN TRIED TO MATCH THE BONE DEFECT. THE BONE GRAFT AND BASE PLATE WERE IMPLANTED AND FIXATED WITH THE BASEPLATE SCREWS. THE REST OF THE REVERS SURGICAL TECHNIQUE WAS FOLLOWED TO COMPLETE THE CASE. PATIENT WAS A (B)(6) MALE, (B)(6). THE FOLLOWING INFORMATION WAS OBTAINED FROM PATIENT MEDICAL RECORDS PROVIDED 3/16/2017: OPERATIVE REPORT (B)(6) 2013 SURGERY: DATE OF ORIGINAL SURGERY WAS (B)(6) 2013. HOSPITAL MEDICAL RECORDS (B)(6) 2017: PATIENT WAS ORIGINALLY SCHEDULED TO BE DISCHARGED ON (B)(6) 2017. SURGEON CHOSE TO EXTEND ONE ADDITIONAL DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214955 ECLIPSE TRUNION,49 MM TPS CTD PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED KWS ARTHREX, INC. 1128006

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other