FDA Adverse Event Injury Summary report: N

BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45MM

MDR report key: 6431582 · Received March 24, 2017

Report

Report Number
8041187-2017-00033
Event Type
Injury
Date Received
March 24, 2017
Date of Event
December 1, 2016
Report Date
April 25, 2017
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Additional Manufacturer Narrative · 1

RESULTS: ONE USED SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL/MICROSCOPIC INSPECTION REVEALED A BROKEN CATHETER THAT APPEARED TO HAVE BEEN CUT BY A SHARP OBJECT. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6230127. A MANUFACTURING REVIEW REVEALED NO ABNORMALITIES WITH PREVENTATIVE MAINTENANCE, CALIBRATION, OR EQUIPMENT THAT COULD HAVE INFLUENCED THE CUSTOMER'S REPORTED ISSUE. CONCLUSION: ALTHOUGH THE RETURNED SAMPLE CONFIRMED A BROKEN CATHETER, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. ADDITIONALLY, OUR QUALITY ENGINEER NOTES THAT THERE IS NO PROCESS IN THE MANUFACTURING FACILITY THAT COULD HAVE CAUSED THIS TYPE OF DEFECT AND STATES THAT THE DEFECT MOST LIKELY OCCURRED OUTSIDE OF THE MANUFACTURING SITE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIP OF A BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45 MM BROKE OFF IN A PATIENT'S RIGHT RADIAL ARTERY AS IT WAS BEING REMOVED. THE PATIENT RECEIVED A DUPLEX SCAN AND THE BROKEN CATHETER WAS REMOVED SURGICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214126 BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45MM ARTERIAL CANNULA FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 6230127

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention