BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45MM
Report
- Report Number
- 8041187-2017-00033
- Event Type
- Injury
- Date Received
- March 24, 2017
- Date of Event
- December 1, 2016
- Report Date
- April 25, 2017
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
(B)(6). PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
RESULTS: ONE USED SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL/MICROSCOPIC INSPECTION REVEALED A BROKEN CATHETER THAT APPEARED TO HAVE BEEN CUT BY A SHARP OBJECT. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6230127. A MANUFACTURING REVIEW REVEALED NO ABNORMALITIES WITH PREVENTATIVE MAINTENANCE, CALIBRATION, OR EQUIPMENT THAT COULD HAVE INFLUENCED THE CUSTOMER'S REPORTED ISSUE. CONCLUSION: ALTHOUGH THE RETURNED SAMPLE CONFIRMED A BROKEN CATHETER, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. ADDITIONALLY, OUR QUALITY ENGINEER NOTES THAT THERE IS NO PROCESS IN THE MANUFACTURING FACILITY THAT COULD HAVE CAUSED THIS TYPE OF DEFECT AND STATES THAT THE DEFECT MOST LIKELY OCCURRED OUTSIDE OF THE MANUFACTURING SITE.
IT WAS REPORTED THAT THE TIP OF A BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45 MM BROKE OFF IN A PATIENT'S RIGHT RADIAL ARTERY AS IT WAS BEING REMOVED. THE PATIENT RECEIVED A DUPLEX SCAN AND THE BROKEN CATHETER WAS REMOVED SURGICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214126 | BD¿ ARTERIAL CANNULA WITH BD FLOWSWITCH¿ 20G X 45MM | ARTERIAL CANNULA | FOZ | BECTON DICKINSON MEDICAL (SINGAPORE) | 6230127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |