IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2017-00093
- Event Type
- Malfunction
- Date Received
- March 23, 2017
- Date of Event
- February 27, 2017
- Report Date
- February 27, 2017
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: -- INDIGO HANDPIECE 1845000 ¿ PRE-REPAIR TEMPERATURES RECORDED AFTER RUNNING THE DEVICE FOR 1 MINUTE AT 60,000 RPM: T1 ¿ 29.6 C, T2 ¿ 26.9 C. BEARINGS WERE REPLACED ON THE DEVICE DUE TO CORROSION. THE DEVICE WAS CLEANED, REPAIRED AND SUCCESSFULLY TESTED TO SPECIFICATIONS. -- STRAIGHT ATTACHMENT 1845010 ¿ PRE-REPAIR TEMPERATURE RECORDED AFTER RUNNING THE DEVICE FOR 1 MINUTE AT 60,000 RPM: 26.3 C. -- ANGLED ATTACHMENT 1845020 ¿ PRE-REPAIR TEMPERATURE RECORDED AFTER RUNNING THE DEVICE FOR 1 MINUTE AT 60,000 RPM: 27.3 C. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
CONCOMITANT DEVICES: 1845010: ATTACHMENT 1845010 INDIGO OTOL STRAIGHT, S/N (B)(4), LOT 206363140, UDI # (B)(4), MANUFACTURED: 2012/11/20 - 1845020: ATTACHMENT 1845020 INDIGO OTOL ANGLED, S/N (B)(4), LOT 206374943, UDI # (B)(4), MANUFACTURED: 2012/11/27. PRODUCT EVALUATION: ANALYSIS RESULTS ARE NOT AVAILABLE; THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING A MASTOIDECTOMY, THE PRODUCT OVERHEATED IMMEDIATELY AND IT WAS TOO HOT TO BE HELD BY THE PHYSICIAN." IT WAS CONFIRMED THAT THE DRILL OVERHEATING WHEN USING EACH OF THE ATTACHMENTS. A BACKUP DRILL SYSTEM WAS USED TO COMPLETE THE PROCEDURE; THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 211224 | IPC® HANDPIECE - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED INC. | 1845000 | 206566959 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |