SYSTEM 98
Report
- Report Number
- 2249723-2017-00008
- Event Type
- Death
- Date Received
- March 22, 2017
- Date of Event
- January 21, 2017
- Report Date
- March 7, 2017
- Manufacturer
- DATASCOPE MAHWAH
- Product Code
- DSP
- PMA / PMN Number
- K961509
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
CORRECTION. IT WAS LATER REPORTED BY THE COMPANY REPRESENTATIVE THAT THE CUSTOMER REPLACED THE FILTER 0008-00-0331 ON (B)(6) 2017 DUE TO BLOOD CONTAMINATION PRIOR TO MAQUET BECOMING AWARE OF THE EVENT ON FEBRUARY 22, 2017. THE PRODUCTION DEVICE HISTORY RECORD (DHR) FOR THIS IABP UNIT WAS REVIEWED AND NO NON-CONFORMANCES RELATED TO THE REPORTED EVENT WERE NOTED. A COMPANY REPRESENTATIVE EVALUATED THE UNIT AND REPLACED THE FILTER (PART NUMBER 0008-00-0331) DUE TO BLOOD CONTAMINATION. THE COMPANY REPRESENTATIVE REPORTED THAT THE UNIT WORKED NORMALLY AT THE TIME OF THE EVENT AND THERE WAS NO MALFUNCTION. THIS EVENT IS NOT INDICATIVE OF A FAILURE OR MALFUNCTION OF THE PUMP BUT SINCE A MAQUET BALLOON WAS INVOLVED AND MALFUNCTIONED AND THE PATIENT DIED WE ARE REPORTING THIS EVENT.
OUR BUSINESS UNIT IN (B)(4) REPORTED THAT AN IAB WAS INSERTED ON (B)(6) 2017 BECAUSE THE PATIENT WAS AMI. THE UNIT WORKED NORMALLY UNTIL (B)(6) 2017 WHEN THE PUMP ALARMED "BLOOD DETECTED" AND BLOOD WAS FOUND IN THE TUBING. THE CUSTOMER REMOVED THE BALLOON AND BECAUSE THE PATIENT WAS OK AT THAT TIME THE DOCTOR NOT INSERT A SECOND BALLOON. A DAY LATER THE PATIENT DIED. THE PATIENT'S DEATH IS NOT ATTRIBUTED TO THE DEVICE. FURTHERMORE, AN ADDITIONAL REPORT IS BEING SUBMITTED WITH THE INFORMATION REGARDING THE INTRA-AORTIC BALLOON ((B)(4)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 208300 | SYSTEM 98 | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE MAHWAH | 0998-00-0446-53 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Death | LINEAR 7.5FR 40CC IAB |