FDA Adverse Event Malfunction Summary report: N

1/2 ML BD SAFETYGLIDE¿ INSULIN SYRINGE WITH 29 G X 1/2 IN. BD¿ NEEDLE

MDR report key: 6424886 · Received March 22, 2017

Report

Report Number
1920898-2017-00037
Event Type
Malfunction
Date Received
March 22, 2017
Date of Event
February 16, 2017
Report Date
April 24, 2017
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K941657
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: ONE USED SAMPLE IN AN OPEN BLISTER PACK WAS RETURNED FOR EVALUATION. A VISUAL INSPECTION REVEALED THAT THE NEEDLE/HUB/SAFETY MECHANISM ASSEMBLY WAS SEPARATED FROM THE BARREL. NO DAMAGE TO THE BARREL WAS OBSERVED. THE RETURNED UNIT WAS FORWARDED TO THE MANUFACTURING SITE IN (B)(6), FOR FURTHER EVALUATION. CONCLUSION: A CONCLUSION IS NOT YET AVAILABLE AS THE SAMPLE IS STILL BEING EVALUATED. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Additional Manufacturer Narrative · 1

FURTHER EVALUATION BY THE MANUFACTURING SITE IN (B)(4) SHOWED THAT A REVIEW OF THE DEVICE HISTORY RECORD REVEALED ZERO DEFECTS NOTED DURING THE MANUFACTURING OF LOT # 6232682. OUR QUALITY ENGINEER NOTES THAT THE PROBABLE CAUSE FOR THE SEPARATED SAFETY MECHANISM IS THAT THERE WAS A PLUNGER OR OTHER SYRINGE COMPONENT CAUGHT IN THE MAIN DIAL OF THE ASSEMBLY EQUIPMENT. THIS PREVENTED THE SPRING FROM FULLY SNAPPING THE ADAPTER/PUSH ROD TO THE HUB.

Description of Event or Problem · 1

IT WAS REPORTED THAT AS A NURSE ACTIVATED THE SAFETY MECHANISM OF A 1/2 ML BD SAFETYGLIDE¿ INSULIN SYRINGE WITH 29 G X 1/2 IN. BD¿ PERMANENTLY ATTACHED NEEDLE, THE NEEDLE AND SAFETY MECHANISM FLEW OFF OF THE SYRINGE. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.