FDA Adverse Event Injury Summary report: N

CODMAN VPV SYSTEM

MDR report key: 6419290 · Received March 20, 2017

Report

Report Number
1226348-2017-10173
Event Type
Injury
Date Received
March 20, 2017
Date of Event
March 1, 2017
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
K061876
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED FOR EVALUATION. UPON COMPLETION OF INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). BOTH 510(K) #'S: K061876 & K050739. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE PRODUCT WAS RECEIVED IN (B)(4) ON MARCH 21ST, 2017. THE PRODUCT WAS TESTED IN CODMAN (B)(4) AND IT WAS REPORTED THAT AN ERROR MESSAGE DISPLAYED. THE ISSUE WAS CONFIRMED. THE DEVICE WAS FORWARDED TO OUR SUPPLIER (B)(4): THE INVESTIGATION AT THE SUPPLIER DETECTED THAT THE FRONT PANEL WAS DAMAGED, THE TEMPERATURE SENSOR HAS NOT BEEN VERIFIED AND THE ATTACHÉ CASE WAS DAMAGED. THE FRONT PANEL AND THE ATTACHÉ CASE WERE REPLACED AND THE TEMPERATURE SENSOR WAS VERIFIED. THE DEVICE WILL BE SHIPPED BACK. AS THE ISSUE WAS RELATING TO A USER ERROR, NO DHR REVIEW WAS PERFORMED FOR THE VPV PROGRAMMER 82-3192, SN (B)(4)(LOT # CPMBK4). BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Description of Event or Problem · 1

DURING SHUNT REPROGRAMMING, ERROR MESSAGE CAME UP SAYING THE TRANSMITTER IS BROKEN. AS A RESULT, PATIENT HAS BEEN ADMITTED TO THE HOSPITAL WHILE A VPV PROGRAMMER IS SOURCED FROM ANOTHER HOSPITAL. PROCEDURE HAS BEEN RE-SCHEDULED ON (B)(6) 2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200281 CODMAN VPV SYSTEM VALVE PROGRAMMER JXG CODMAN & SHURTLEFF, INC. / MEDOS S.A. CPMBK4

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention