FDA Adverse Event Malfunction Summary report: N

ELECSYS FT3 III

MDR report key: 6419150 · Received March 20, 2017

Report

Report Number
1823260-2017-00592
Event Type
Malfunction
Date Received
March 20, 2017
Date of Event
January 31, 2017
Report Date
April 6, 2017
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CDP
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SAMPLE FROM THE PATIENT WAS PROVIDED FOR INVESTIGATION. INVESTIGATIONS HAVE DETERMINED THAT THE SAMPLE CONTAINS AN INTERFERENT TO COMPONENTS USED IN THE FT3 AND FT4 ASSAYS. THIS LIMITATION IS COVERED IN PRODUCT LABELING.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED IN (B)(6). UNIQUE IDENTIFIER (UDI)#: (B)(4). FACILITY NAME - THE FULL FACILITY NAME WAS PROVIDED AS "(B)(6)".

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THEY RECEIVED ERRONEOUS RESULTS FOR TWO SAMPLES FROM THE SAME PATIENT TESTED FOR THE ELECSYS FT3 III (FT3) AND THE ELECSYS FT4 II ASSAY (FT4) ON A COBAS 8000 E 602 MODULE. THE RESULTS WERE SAID THE BE OUTSIDE OF THE RESPECTIVE REFERENCE RANGES FOR THE ROCHE ASSAYS, BUT WERE NORMAL WHEN TESTED WITH THE SAME METHODS ON AN ABBOTT ANALYZER. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE PATIENT DOES NOT HAVE ANY SYMPTOMS OF THYROID DISEASE. THIS MEDWATCH WILL COVER FT3 REAGENT LOT NUMBER 179026. PLEASE REFER TO THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO FT3 REAGENT LOT NUMBER 152267, REFER TO THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO FT4 REAGENT LOT NUMBER 158223, AND REFER TO THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO FT4 REAGENT LOT NUMBER 180429. PLEASE REFER TO THE ATTACHMENT FOR ALL PATIENT DATA. THE SAMPLE FROM (B)(6) 2016 WAS TESTED WITH FT3 REAGENT LOT 152267 AND FT4 REAGENT LOT 158223. THE SAMPLE FROM (B)(6) 2017 WAS TESTED WITH FT3 REAGENT LOT 179026 AND FT4 REAGENT LOT 180429. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE E602 ANALYZER SERIAL NUMBER WAS ASKED FOR, BUT NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201041 ELECSYS FT3 III RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE CDP ROCHE DIAGNOSTICS NA 179026

Patients

Seq Age Sex Outcome Treatment
1 APDIRA| LANTUS