FDA Adverse Event Malfunction Summary report: N

ORCHESTRA

MDR report key: 6416619 · Received March 20, 2017

Report

Report Number
1000165971-2017-00225
Event Type
Malfunction
Date Received
March 20, 2017
Date of Event
February 7, 2017
Report Date
February 20, 2017
Manufacturer
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
Product Code
DXY
PMA / PMN Number
P980049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY ANALYSIS SHOWED THAT THE DISPLAYED EGM WAS MOST PROBABLY CHANGED MANUALLY BY THE USER FROM EGM A TO EGM V BEFORE LAUNCHING THE ATRIAL THRESHOLD TEST. THE DISPLAYED CHANNEL REMAINS THE ONE MANUALLY SELECTED BY THE USER, AS EXPECTED.

Description of Event or Problem · 1

REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.

Description of Event or Problem · 1

REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.

Description of Event or Problem · 1

REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199066 ORCHESTRA PROGRAMMER, PACEMAKER DXY SORIN GROUP ITALIA S.R.L. - CRM FACILITY ORCHESTRA PLUS

Patients

Seq Age Sex Outcome Treatment
1