ORCHESTRA
Report
- Report Number
- 1000165971-2017-00225
- Event Type
- Malfunction
- Date Received
- March 20, 2017
- Date of Event
- February 7, 2017
- Report Date
- February 20, 2017
- Manufacturer
- SORIN GROUP ITALIA S.R.L. - CRM FACILITY
- Product Code
- DXY
- PMA / PMN Number
- P980049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
PRELIMINARY ANALYSIS SHOWED THAT THE DISPLAYED EGM WAS MOST PROBABLY CHANGED MANUALLY BY THE USER FROM EGM A TO EGM V BEFORE LAUNCHING THE ATRIAL THRESHOLD TEST. THE DISPLAYED CHANNEL REMAINS THE ONE MANUALLY SELECTED BY THE USER, AS EXPECTED.
REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.
REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.
REPORTEDLY, DURING A PACEMAKER FOLLOW-UP PERFORMED ON (B)(6) 2017, IT WAS OBSERVED THAT WHEN ¿A LEAD¿ WAS SELECTED FOR THRESHOLD TESTING, THE DISPLAYED EGM DID NOT AUTOMATICALLY CHANGE FROM EGM V TO EGM A. AFTERWARDS, THE OPERATOR MANUALLY HAD EGM A DISPLAYED AND STARTED A THRESHOLD TEST AGAIN. ATRIAL THRESHOLD TESTING WAS THEN SUCCESSFULLY PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 199066 | ORCHESTRA | PROGRAMMER, PACEMAKER | DXY | SORIN GROUP ITALIA S.R.L. - CRM FACILITY | ORCHESTRA PLUS |