FDA Adverse Event Malfunction Summary report: N

HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°, STERILE, SINGLE USE

MDR report key: 6412841 · Received March 17, 2017

Report

Report Number
9610773-2017-00045
Event Type
Malfunction
Date Received
March 17, 2017
Date of Event
March 9, 2017
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FAS
PMA / PMN Number
PK931763
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THREE HF RESECTION ELECTRODES FROM THE SAME MODEL AND LOT WERE RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION. THE VISUAL INSPECTION CONFIRMED THAT THE LOOP WIRES AT THE DISTAL END OF TWO OF THE HF RESECTION ELECTRODES ARE BROKEN OFF. ONE OF THE HF RESECTION ELECTRODES IS UNUSED, IN ITS ORIGINAL STERILE PACKAGING. OLYMPUS AMERICA INC. ("OAI") IS IMPLEMENTING A REMOVAL ACTION OF SPECIFIED LOT NUMBERS OF THE MONOPOLAR HF RESECTION ELECTRODES A22201C AND WA22037C. THE ELECTRODES ARE USED FOR ENDOSCOPIC DIAGNOSIS AND TREATMENT IN UROLOGICAL AND GYNECOLOGICAL APPLICATIONS. OAI HAS INITIATED THIS REMOVAL ACTION AFTER RECEIVING AN INCREASED NUMBER OF COMPLAINTS REGARDING LOOP WIRES BREAKING AT THE DISTAL END OF THE REFERENCED ELECTRODES. INVESTIGATIONS HAVE CONFIRMED THAT LOOP WIRES CAN BREAK DURING THE INTENDED USE OF THE ELECTRODES. AS A RESULT, A FRAGMENT MAY FALL INSIDE THE PATIENT AND WILL NEED TO BE RETRIEVED. RETRIEVAL OF THE FRAGMENT COULD PROLONG THE PROCEDURE AND, UNDER CERTAIN CIRCUMSTANCES, COULD REQUIRE ADDITIONAL SURGICAL TREATMENT. THE INVESTIGATION REVEALED THAT THE LOOP WIRES OF THE AFFECTED ELECTRODES WERE DAMAGED DURING PRODUCTION. THE CAUSE OF THIS DAMAGE IS DEFECTIVE MANUFACTURING EQUIPMENT. THE DAMAGED LOOP WIRES CANNOT BE DETECTED BY VISUAL INSPECTION. THERE HAS BEEN NO REPORT TO DATE OF AN ADVERSE EVENT OR PATIENT INJURY. HOWEVER, IN AN EFFORT TO PREVENT A POTENTIAL RISK TO PATIENT HEALTH, OAI IS UNDERTAKING THIS ACTION TO REMOVE THE AFFECTED LOT NUMBERS. OAI'S CORRECTION NUMBER ACCORDING TO 21 CFR 806.10 (C) (1): 2429304-4/18/2017-044R

Additional Manufacturer Narrative · 1

THE SUSPECT MEDICAL DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE USER'S EXPERIENCE AND THE REPORTED PHENOMENON COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, IF THE SUSPECT MEDICAL DEVICE IS RETURNED FOR EVALUATION/INVESTIGATION OR ADDITIONAL SIGNIFICANT INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED THERAPEUTIC TRANSURETHRAL RESECTION (TUR) PROCEDURE, THE LOOP WIRE AT THE DISTAL END OF THE HF RESECTION ELECTRODE BROKE OFF AND FELL INSIDE THE PATIENT. HOWEVER, NO FRAGMENT REMAINED INSIDE THE PATIENT SINCE IT WAS REPORTEDLY RETRIEVED BY UNKNOWN APPROACH. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER HF RESECTION ELECTRODE AND THERE WAS NO ADVERSE EVENT OR PATIENT INJURY. THIS IS REPORT 1 OUT OF 3.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197616 HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°, STERILE, SINGLE USE HF-RESECTION ELECTRODES FAS OLYMPUS WINTER & IBE GMBH A22201C P16Z0001

Patients

Seq Age Sex Outcome Treatment
1 UNSPECIFIED VALLEYLAB ELECTROSURGICAL GENERATOR