FDA Adverse Event Malfunction Summary report: N

DIETER MALLEUS NIPPER UP-CUT 75MM

MDR report key: 6407542 · Received March 15, 2017

Report

Report Number
9610612-2017-00092
Event Type
Malfunction
Date Received
March 15, 2017
Date of Event
February 3, 2017
Report Date
December 5, 2017
Manufacturer
AESCULAP AG
Product Code
JYR
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF AESCULAP AG (MANUFACTURER). EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Description of Event or Problem · 0

COUNTRY OF COMPLAINT: UNITED KINGDOM. IT WAS REPORTED THAT ONE OF THE INSTRUMENTS (DIETER MALLEUS NIPPE) BROKE. IT WAS REPLACED WHICH CAUSE A 20-30 MINUTE DELAY IN SURGERY. THE BROKEN PIECE WAS RETRIEVED. THERE WAS NO HARM TO THE PATIENT REPORTED.

Additional Manufacturer Narrative · 1

INVESTIGATION: THE INVESTIGATION WAS PERFORMED USING A KEYENCE VHX-5000 DIGITAL MICROSCOPE. THE ANALYSIS OF THE FRACTURE PATTERN ILLUSTRATED A FORCED FRACTURE DUE TO OVERLOAD. NO PORES, INCLUSIONS OR FOREIGN BODIES COULD BE FOUND ON THE POINT OF RUPTURE. BATCH HISTORY REVIEW: THE DEVICE QUALITY AND MANUFACTURING HISTORY RECORDS HAVE BEEN CHECKED FOR ALL AVAILABLE LOT NUMBERS. THE DEVICE HISTORY FILE HAS BEEN CHECKED AND FOUND TO BE ACCORDING TO OUR SPECIFICATION VALID AT THE TIME OF PRODUCTION. NO SIMILAR INCIDENTS HAVE BEEN FILED WITH PRODUCTS FROM THIS BATCH/THESE BATCHES. CONCLUSION AND ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF OUR INVESTIGATION THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. RATIONAL: THIS KIND OF INSTRUMENT IS DESIGNED FOR DELICATE USE ONLY. IT IS LIABLE THAT A MECHANICAL OVERLOAD SITUATION LED TO THE BREAKAGE. THE VISUAL INVESTIGATION INDICATED THAT THE QUALITY REQUIREMENTS ARE IN THE SPECIFIED TOLERANCE RANGE. NO CAPA IS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188063 DIETER MALLEUS NIPPER UP-CUT 75MM GENERAL SURGICAL INSTRUMENTS JYR AESCULAP AG OG300R

Patients

Seq Age Sex Outcome Treatment
1 Other