REAGENT RED BLOOD CELLS 0.8% RESOLVE PANEL A
Report
- Report Number
- 2250051-2017-00018
- Event Type
- Malfunction
- Date Received
- March 15, 2017
- Date of Event
- February 27, 2017
- Report Date
- March 15, 2017
- Manufacturer
- ORTHO CLINICAL DIAGNOSTICS
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ORTHO PERFORMED RETAIN TESTING, BATCH REVIEW, COMPLAINT REVIEW BY LOT, DONOR HISTORY, AND DONOR COMPLAINT REVIEW. ALL RESULTS WERE SATISFACTORY. SAMPLE WAS NOT RETURNED TO ORTHO FOR FURTHER INVESTIGATION. (B)(4).
EVENT 1 OF 4. CUSTOMER REPORTED FALSE NEGATIVES WITH 4 SEPARATE PATIENTS IN MANUAL GEL METHOD WITH PANEL CELL NUMBER# 6 OF 0.8% RESOLVE PANEL A LOT# VRA270 USING MTS IGG GEL CARD. CUSTOMER STATED THE FOUR PATIENTS HAD PREVIOUS HISTORY OF ANTI-E. CUSTOMER RAN THE PANELS BECAUSE IT IS PART OF THEIR PROTOCOL TO RUN A PANEL ON PATIENTS WITH PREVIOUS HISTORY OF AN ANTIBODY. ALL FOUR PATIENTS REACTED WITH PANEL CELL #3 (HOMOZYGOUS FOR E) WITH 2+ REACTIVITY. ISSUE STARTED ON: (B)(6) 2017; REPORTED 3-02-2017. FREQUENCY: 4X. (B)(6). OTHER RELEVANT DETAILS: ALL TESTING PERFORMED USING GEL CARDS LOT # 101016001-21. CUSTOMER RAN RESOLVE PANEL B SELECTED CELLS FOR IDENTIFICATION AND RULE OUT. ORTHO CARE ASKED THE CUSTOMER TO TEST WITH AN ALTERNATE LOT OF IGG GEL CARDS, BUT CUSTOMER DID NOT HAVE AN ALTERNATE LOT. ORTHO CARE ASKED CUSTOMER IF RED CELL REAGENTS WERE ALLOWED TO COME TO ROOM TEMPERATURE, CUSTOMER REPORTS YES. ORTHO CARE ASKED THE CUSTOMER TO TEST PANEL CELL #6 WITH ANTI-E REAGENT. ORTHO CARE FOLLOWED UP WITH THE CUSTOMER TO CHECK THE RESULTS OF TESTING USING ANTI-E ANTISERA WITH PANEL CELL #6. CUSTOMER STATED THAT PANEL CELL #6 HAD 2+ REACTIVITY WITH ANTISERA. ORTHO CARE DISCUSSED WITH THE CUSTOMER THAT THE ISSUE IS SAMPLE RELATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 188037 | REAGENT RED BLOOD CELLS 0.8% RESOLVE PANEL A | REAGENT RED BLOOD CELLS | KSZ | ORTHO CLINICAL DIAGNOSTICS | VRA270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |