FDA Adverse Event Injury Summary report: N

COMPR SRS IC SEG - 60MM

MDR report key: 6404038 · Received March 14, 2017

Report

Report Number
0001825034-2017-01566
Event Type
Injury
Date Received
March 14, 2017
Date of Event
February 6, 2017
Report Date
November 15, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWT
PMA / PMN Number
PK111746
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT X RAY¿S WERE EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. AN X-RAY WAS PROVIDED AND REVIEWED. THE RESULTS OF THE REVIEW IDENTIFIED THE FOLLOWING: "COMPREHENSIVE SEGMENTAL DISTAL HUMERAL/ELBOW RECONSTRUCTION PROSTHESES ARE PRESENT. THE HUMERAL DISTAL BODY COMPONENT IS COMPLETELY DISSOCIATED FROM THE HUMERAL STEM COMPONENT, AND IS DISPLACED MEDIALLY. A STABILIZING PLATE AND SCREW ARE NOT IDENTIFIED IN THE VICINITY OF JUNCTION OF HUMERAL STEM AND BODY COMPONENTS." ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

(B)(4). MEDICAL PRODUCT - COMPR SRS ANTI ROT IC SEG-30MM, CAT#: 211266 LOT#: 515410; COMPR SRS MOD STEM - 10X100MM, CAT#: 211237 LOT#: 871110; HUMERAL SCREW KIT 2 HUMERAL SCREWS, CAT#: 00840009000 LOT#: 63425387; ARTICULATION KIT SIZE 5/6 1 AXLE PIN, 1 HUMERAL BEARING A, 2 ULNAR BEARINGS B, CAT#: 00840009500 LOT#: 63396483; COMPR SRS LARGE FLANGE, CAT#: 211270 LOT#: 231150; ELBOW TORQUE DRIVER, CAT#: 00840108000 LOT#: 56613805; COMPR SRS 60MM DST HUM BDY LT, CAT#: 110029825 LOT#: 115510. THERAPY DATE ¿ UNKNOWN DATE IN (B)(6) 2017. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED [AS IT REMAINS IMPLANTED AT THE TIME OF THIS REPORT] TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS:1825034-2017-01558 / 01566.

Description of Event or Problem · 1

IT HAS BEEN REPORTED THAT THE PATIENT UNDERWENT A LEFT ELBOW ARTHROPLASTY AND APPROXIMATELY 12 DAYS POST-IMPLANTATION THE PATIENT STATED HEARING A CLICKING SOUND. RADIOGRAPHS REVEAL THAT THE HUMERAL COMPONENT HAS FAILED AND SEPARATED FROM THE INTERCALARY SEGMENT. NO REVISION HAS BEEN REPORTED TO DATE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186353 COMPR SRS IC SEG - 60MM PROSTHESIS, SHOULDER KWT BIOMET ORTHOPEDICS N/A 104420

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention SEE NARRATIVE IN H10