FDA Adverse Event Malfunction Summary report: N

HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°

MDR report key: 6402059 · Received March 14, 2017

Report

Report Number
9610773-2017-00038
Event Type
Malfunction
Date Received
March 14, 2017
Date of Event
March 9, 2017
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCP
PMA / PMN Number
PK790071
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE SUSPECT MEDICAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION. THE VISUAL INSPECTION CONFIRMED THAT THE LOOP WIRE AT THE DISTAL END OF THE HF RESECTION ELECTRODE IS BROKEN OFF. OLYMPUS AMERICA INC. ("OAI") IS IMPLEMENTING A REMOVAL ACTION OF SPECIFIED LOT NUMBERS OF THE MONOPOLAR HF RESECTION ELECTRODES A22201C AND WA22037C. THE ELECTRODES ARE USED FOR ENDOSCOPIC DIAGNOSIS AND TREATMENT IN UROLOGICAL AND GYNECOLOGICAL APPLICATIONS. OAI HAS INITIATED THIS REMOVAL ACTION AFTER RECEIVING AN INCREASED NUMBER OF COMPLAINTS REGARDING LOOP WIRES BREAKING AT THE DISTAL END OF THE REFERENCED ELECTRODES. INVESTIGATIONS HAVE CONFIRMED THAT LOOP WIRES CAN BREAK DURING THE INTENDED USE OF THE ELECTRODES. AS A RESULT, A FRAGMENT MAY FALL INSIDE THE PATIENT AND WILL NEED TO BE RETRIEVED. RETRIEVAL OF THE FRAGMENT COULD PROLONG THE PROCEDURE AND, UNDER CERTAIN CIRCUMSTANCES, COULD REQUIRE ADDITIONAL SURGICAL TREATMENT. THE INVESTIGATION REVEALED THAT THE LOOP WIRES OF THE AFFECTED ELECTRODES WERE DAMAGED DURING PRODUCTION. THE CAUSE OF THIS DAMAGE IS DEFECTIVE MANUFACTURING EQUIPMENT. THE DAMAGED LOOP WIRES CANNOT BE DETECTED BY VISUAL INSPECTION. THERE HAS BEEN NO REPORT TO DATE OF AN ADVERSE EVENT OR PATIENT INJURY. HOWEVER, IN AN EFFORT TO PREVENT A POTENTIAL RISK TO PATIENT HEALTH, OAI IS UNDERTAKING THIS ACTION TO REMOVE THE AFFECTED LOT NUMBERS. OAI'S CORRECTION NUMBER ACCORDING TO 21 CFR 806.10 (C) (1): 2429304-4/18/2017-044R

Additional Manufacturer Narrative · 1

THE SUSPECT MEDICAL DEVICES HAVE NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE USER'S EXPERIENCE AND THE REPORTED PHENOMENON COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, IF THE SUSPECT MEDICAL DEVICES ARE RETURNED FOR EVALUATION/INVESTIGATION OR ADDITIONAL SIGNIFICANT INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE BLADDER (TURB) OR PROSTATE (TURP) PROCEDURE, THE LOOP WIRE AT THE DISTAL END OF AT LEAST ONE HF RESECTION ELECTRODE BROKE OFF AND FELL INSIDE THE PATIENT. HOWEVER, NO FRAGMENTS REMAINED INSIDE THE PATIENT SINCE THEY WERE REPORTEDLY RETRIEVED BY UNKNOWN APPROACH. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER HF RESECTION ELECTRODE AND THERE WAS NO ADVERSE EVENT OR PATIENT INJURY. THIS IS REPORT 2 OUT OF 3.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184834 HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12° HF-RESECTION ELECTRODES GCP OLYMPUS WINTER & IBE GMBH A22201A P16X0002

Patients

Seq Age Sex Outcome Treatment
1