FDA Adverse Event Malfunction Summary report: N

CODMAN VPV SYSTEM

MDR report key: 6399007 · Received March 13, 2017

Report

Report Number
1226348-2017-10151
Event Type
Malfunction
Date Received
March 13, 2017
Date of Event
February 20, 2017
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
K061876
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BOTH 510(K) #'S: K061876 & K050739. UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

UDI - (B)(4). THE PRODUCT WAS TESTED IN CODMAN (B)(4) AND IT WAS REPORTED THAT THE MESSAGE ¿REPEAT ADJUSTMENT¿ DISPLAYED. THE ISSUE WAS CONFIRMED. THE DEVICE WAS FORWARDED TO THE SUPPLIER. THE INVESTIGATION AT THE SUPPLIER DETECTED THAT THERE WAS A BAD CONNECTION IN THE SENSING UNIT. THE TRANSMITTER CABLE WAS REPLACED. THE DEVICE WILL BE SHIPPED BACK TO THE CUSTOMER. A DHR REVIEW WAS PERFORMED FOR THE PROGRAMMER 82-3192 S/N: (B)(4) (LOT# CJPBDV), AND THE LOT MET SPECIFICATIONS WHEN RELEASED ON DECEMBER 11TH, 2008. TRENDS WILL BE MONITORED FOR THIS OR SIMILAR COMPLAINTS. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.

Description of Event or Problem · 1

THE CLINICIAN CONTACTED ME TO REPORT THAT HE WAS TRYING TO PROGRAM A VALVE HE HAD RECENTLY IMPLANTED INTO A PATIENT USING THE LOAN VPV PROGRAMMER CONSIGNED TO HIS ROOMS, HOWEVER HE WAS NOT HAVING ANY LUCK AS THE TRANSMITTER HEAD ON THE VPV PROGRAMMER WAS HEATING UP AND FLASHING AN ERROR MESSAGE. REP VISITED THE CLINICIAN'S ROOMS YESTERDAY TO DROP OFF A WORKING LOAN VPV PROGRAMMER AND TO TEST IF THE TRANSMITTER HEAD WAS HEATING UP AND THE REP CAN CONFIRM THAT THIS IS TRUE. PT WAS UNABLE TO HAVE THEIR VALVE ADJUSTED UNTIL 1 DAY LATER, WHEN REP DROPPED OFF A LOAN UNIT.