CODMAN VPV SYSTEM
Report
- Report Number
- 1226348-2017-10151
- Event Type
- Malfunction
- Date Received
- March 13, 2017
- Date of Event
- February 20, 2017
- Manufacturer
- CODMAN & SHURTLEFF, INC. / MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K061876
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). BOTH 510(K) #'S: K061876 & K050739. UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.
UDI - (B)(4). THE PRODUCT WAS TESTED IN CODMAN (B)(4) AND IT WAS REPORTED THAT THE MESSAGE ¿REPEAT ADJUSTMENT¿ DISPLAYED. THE ISSUE WAS CONFIRMED. THE DEVICE WAS FORWARDED TO THE SUPPLIER. THE INVESTIGATION AT THE SUPPLIER DETECTED THAT THERE WAS A BAD CONNECTION IN THE SENSING UNIT. THE TRANSMITTER CABLE WAS REPLACED. THE DEVICE WILL BE SHIPPED BACK TO THE CUSTOMER. A DHR REVIEW WAS PERFORMED FOR THE PROGRAMMER 82-3192 S/N: (B)(4) (LOT# CJPBDV), AND THE LOT MET SPECIFICATIONS WHEN RELEASED ON DECEMBER 11TH, 2008. TRENDS WILL BE MONITORED FOR THIS OR SIMILAR COMPLAINTS. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
THE CLINICIAN CONTACTED ME TO REPORT THAT HE WAS TRYING TO PROGRAM A VALVE HE HAD RECENTLY IMPLANTED INTO A PATIENT USING THE LOAN VPV PROGRAMMER CONSIGNED TO HIS ROOMS, HOWEVER HE WAS NOT HAVING ANY LUCK AS THE TRANSMITTER HEAD ON THE VPV PROGRAMMER WAS HEATING UP AND FLASHING AN ERROR MESSAGE. REP VISITED THE CLINICIAN'S ROOMS YESTERDAY TO DROP OFF A WORKING LOAN VPV PROGRAMMER AND TO TEST IF THE TRANSMITTER HEAD WAS HEATING UP AND THE REP CAN CONFIRM THAT THIS IS TRUE. PT WAS UNABLE TO HAVE THEIR VALVE ADJUSTED UNTIL 1 DAY LATER, WHEN REP DROPPED OFF A LOAN UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182341 | CODMAN VPV SYSTEM | VALVE PROGRAMMER | JXG | CODMAN & SHURTLEFF, INC. / MEDOS S.A. | CJPBDV |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |