FDA Adverse Event Malfunction Summary report: N

RESTORE ADVANCED

MDR report key: 6380341 · Received March 6, 2017

Report

Report Number
3004209178-2017-05063
Event Type
Malfunction
Date Received
March 6, 2017
Date of Event
December 30, 2013
Report Date
May 10, 2017
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) OF A CLINICAL STUDY REPORTED THAT THE SIGNS AND SYMPTOMS INCLUDED COMPLAINTS OF INCREASED SPASMS IN UPPER BACK WITH STIMULATION DUE TO INCREASED AMPLITUDE, AMPLITUDE WAS DECREASED AND SYMPTOMS WERE RESOLVED. THE AMPLITUDE PRIOR TO REPROGRAMMING WAS 5.5 IN PROGRAM B AND 3.4 IN PROGRAM J. THE AMPLITUDE WAS CHANGED TO 2.9 IN B AND 1.8 IN J.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL OF A CLINICAL STUDY REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR OTHER NON-MALIGNANT PAIN AND OTHER CHRONIC/INTRACTABLE PAIN OF THE TRUNK/LIMBS. A CLINICAL DIAGNOSIS OF UNSATISFACTORY STIMULATION WAS REPORTED (AND A DEVICE DIAGNOSIS WAS NOT APPLICABLE). ON (B)(6) 2013, THE PATIENT COMPLAINED OF "GRABBING" IN THE UPPER BACK WITH STIMULATION; THERE WERE COMPLAINTS OF INCREASED SPASMS IN THE UPPER BACK WITH STIMULATION. THE ENTIRE SYSTEM WAS REPROGRAMMED AND THE EVENT RESOLVED WITHOUT SEQUELAE. THE EVENT WAS RELATED TO THE DEVICE OR THERAPY, SPECIFICALLY DUE TO PROGRAMMING, AND WAS NOT RELATED TO THE IMPLANT PROCEDURE. THE FINAL PROGRAMMING DATE AND TIME FOR THE MOST RECENT INTERROGATION PRIOR TO THE EVENT OCCURRED ON (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
164791 RESTORE ADVANCED STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37713

Patients

Seq Age Sex Outcome Treatment
1 54 YR