FLEXLAB
Report
- Report Number
- 3010825766-2017-00001
- Event Type
- Malfunction
- Date Received
- March 6, 2017
- Date of Event
- February 6, 2017
- Report Date
- March 6, 2017
- Manufacturer
- INPECO SA
- Product Code
- JQP
- PMA / PMN Number
- K121012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
THE EVENT WAS INVESTIGATED BY INPECO, WHICH DETERMINED THAT THE ABNORMAL NUMBER OF CLOT DETECTION ERRORS AND THE SAMPLES DILUTION WERE DUE TO A LACK OF MAINTENANCE OF THE ALIQUOTER TUBING SYSTEM. THE ALIQUOTER SERVICE MANUAL IS REFERENCED IN THE FLEXLAB SERVICE MANUAL: THE MAINTENANCE INSTRUCTIONS CLEARLY INDICATES THE FREQUENCY OF ALIQUOTER TUBING SYSTEM REPLACEMENT ACCORDING TO THE TIME PASSED SINCE THE INSTALLATION OR THE PREVIOUS REPLACEMENT OR THE OPERATING HOURS. THE MALFUNCTION WAS NOT DUE TO THE ALIQUOTER MODULE DESIGN. THIS DEVICE IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
THE DISTRIBUTOR SERVICE TEAM HAS BEEN CONTACTED BY THE LABORATORY PERSONNEL DUE TO AN ABNORMAL NUMBER OF CLOT DETECTION ERRORS ON SAMPLES PROCESSED ON ALIQUOTER MODULE #1 (AQM # 1) CONNECTED TO FLEXLAB PRODUCT. DURING THE TROUBLESHOOTING A VIDEO WAS TAKEN TO AID WITH IDENTIFYING THE CAUSE OF THE PROBLEM: IT WAS DETERMINED FROM THE VIDEO THAT WATER FROM THE AQM #1 PIPETTOR CIRCUIT WAS BEING DISPENSED INTO THE SAMPLE TUBES. THE LABORATORY PERSONNEL AND THE DISTRIBUTOR SERVICE TEAM IDENTIFIED 175 PRIMARY SAMPLE TUBES AND 118 SECONDARY SAMPLE TUBES DILUTED DUE TO THE ALIQUOTER MODULE #1 MALFUNCTION. THERE WAS NO DEATH NEITHER SERIOUS INJURY TO ANY PATIENT DUE TO ERRONEOUS TEST RESULTS. AS IMMEDIATE ACTIONS THE ALIQUOTER MODULE #1 WAS SET OFF-LINE AND THE TESTS WERE RE-EXECUTED ON ALTERNATE SAMPLE TUBES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 164882 | FLEXLAB | LABORATORY AUTOMATION SYSTEM | JQP | INPECO SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |