FDA Adverse Event Injury Summary report: N

JUSTRIGHT STAPLER

MDR report key: 6378726 · Received March 4, 2017

Report

Report Number
3010377594-2017-00002
Event Type
Injury
Date Received
March 4, 2017
Date of Event
February 9, 2017
Report Date
March 3, 2017
Manufacturer
JUSTRIGHT SURGICAL, LLC
Product Code
GAG
UDI-DI
10865163000109
PMA / PMN Number
K132472
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO ROOT CAUSE COULD BE DETERMINED BECAUSE THE DEVICE WAS NOT AVAILABLE FOR RETURN OR EVALUATION. THERE WAS EVIDENCE DURING THE PROCEDURE THAT THE DEVICE FUNCTIONED PROPERLY. THE DEVICE WAS DISPOSED OF ON (B)(6) 2017 AFTER THE APPENDECTOMY PROCEDURE WAS PERFORMED. JUST RIGHT SURGICAL REVIEWED THE MANUFACTURING LOT RELEASE INFORMATION AND CONFIRMED THIS LOT PASSED THE FUNCTIONAL ACCEPTANCE CRITERIA AS SPECIFIED IN THE STAPLER AND RELOAD PRODUCTION QUALITY PLAN, INCLUDING STAPLE FORMATION.

Description of Event or Problem · 1

ON (B)(6) 2017, THE SURGEON PERFORMED A LAPAROSCOPICALLY ASSISTED APPENDECTOMY ON A (B)(6) -YEAR OLD PATIENT WEIGHING (B)(4). THE APPENDIX WAS PULLED OUT OF THE UMBILICAL INCISION AND THE STAPLER WAS FIRED EXTRACORPOREALLY. THE STAPLE LINE WAS INSPECTED AND ALL STAPLES WERE PROPERLY FORMED AND CLOSED. THE SURGEON THEN REINSERTED THE APPENDICEAL STUMP INTO THE ABDOMINAL CAVITY THROUGH THE UMBILICUS. THE PATIENT WENT BACK TO THE HOSPITAL ON (B)(6) 2017 FOR A CORRECTIVE PROCEDURE FOR A CLOSED LOOP BOWEL OBSTRUCTION. THE SURGEON STATED THE OBSTRUCTION WAS CAUSED BY A STAPLE THAT HAD BECOME ATTACHED TO PART OF THE BOWEL SEROSA. THE STAPLE WAS REMOVED AND THE PATIENT WAS REPORTED AS FINE. IMAGES FROM THE PROCEDURE INDICATE THAT THE STAPLE WAS FULLY FORMED BUT HAD CAUGHT ON ANOTHER SECTION OF BOWEL FORMING A LOOP. IT IS POSSIBLE THAT THIS OCCURRED WHEN THE STAPLED BOWEL WAS FORCED THROUGH THE UMBILICUS. THE DEVICE WAS DISPOSED OF ON (B)(6) 2017 AND IS NOT AVAILABLE FOR RETURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
161218 JUSTRIGHT STAPLER SURGICAL STAPLER GAG JUSTRIGHT SURGICAL, LLC JR-ST25-2.0-6 75IF1901 10865163000109

Patients

Seq Age Sex Outcome Treatment
1 12 YR Required Intervention