JUSTRIGHT STAPLER
Report
- Report Number
- 3010377594-2017-00002
- Event Type
- Injury
- Date Received
- March 4, 2017
- Date of Event
- February 9, 2017
- Report Date
- March 3, 2017
- Manufacturer
- JUSTRIGHT SURGICAL, LLC
- Product Code
- GAG
- UDI-DI
- 10865163000109
- PMA / PMN Number
- K132472
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
NO ROOT CAUSE COULD BE DETERMINED BECAUSE THE DEVICE WAS NOT AVAILABLE FOR RETURN OR EVALUATION. THERE WAS EVIDENCE DURING THE PROCEDURE THAT THE DEVICE FUNCTIONED PROPERLY. THE DEVICE WAS DISPOSED OF ON (B)(6) 2017 AFTER THE APPENDECTOMY PROCEDURE WAS PERFORMED. JUST RIGHT SURGICAL REVIEWED THE MANUFACTURING LOT RELEASE INFORMATION AND CONFIRMED THIS LOT PASSED THE FUNCTIONAL ACCEPTANCE CRITERIA AS SPECIFIED IN THE STAPLER AND RELOAD PRODUCTION QUALITY PLAN, INCLUDING STAPLE FORMATION.
ON (B)(6) 2017, THE SURGEON PERFORMED A LAPAROSCOPICALLY ASSISTED APPENDECTOMY ON A (B)(6) -YEAR OLD PATIENT WEIGHING (B)(4). THE APPENDIX WAS PULLED OUT OF THE UMBILICAL INCISION AND THE STAPLER WAS FIRED EXTRACORPOREALLY. THE STAPLE LINE WAS INSPECTED AND ALL STAPLES WERE PROPERLY FORMED AND CLOSED. THE SURGEON THEN REINSERTED THE APPENDICEAL STUMP INTO THE ABDOMINAL CAVITY THROUGH THE UMBILICUS. THE PATIENT WENT BACK TO THE HOSPITAL ON (B)(6) 2017 FOR A CORRECTIVE PROCEDURE FOR A CLOSED LOOP BOWEL OBSTRUCTION. THE SURGEON STATED THE OBSTRUCTION WAS CAUSED BY A STAPLE THAT HAD BECOME ATTACHED TO PART OF THE BOWEL SEROSA. THE STAPLE WAS REMOVED AND THE PATIENT WAS REPORTED AS FINE. IMAGES FROM THE PROCEDURE INDICATE THAT THE STAPLE WAS FULLY FORMED BUT HAD CAUGHT ON ANOTHER SECTION OF BOWEL FORMING A LOOP. IT IS POSSIBLE THAT THIS OCCURRED WHEN THE STAPLED BOWEL WAS FORCED THROUGH THE UMBILICUS. THE DEVICE WAS DISPOSED OF ON (B)(6) 2017 AND IS NOT AVAILABLE FOR RETURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 161218 | JUSTRIGHT STAPLER | SURGICAL STAPLER | GAG | JUSTRIGHT SURGICAL, LLC | JR-ST25-2.0-6 | 75IF1901 | 10865163000109 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Required Intervention |