FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 6376895 · Received March 3, 2017

Report

Report Number
1220908-2017-00429
Event Type
Malfunction
Date Received
March 3, 2017
Report Date
February 13, 2017
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED INTERCONNECTION FLEX CABLE. THE FLEX CABLE WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "PACER DEFIB DEVICE FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
158263 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1