FDA Adverse Event Injury Summary report: N

AMISTEM H, HA COATED STD SIZE 4 STEM

MDR report key: 6376239 · Received March 3, 2017

Report

Report Number
3005180920-2017-00091
Event Type
Injury
Date Received
March 3, 2017
Date of Event
February 2, 2017
Report Date
March 3, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 17 FEBRUARY 2017 THE MEDICAL AFFAIRS DIRECTOR PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: EARLY POSTOPERATIVE FRACTURE AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD PATIENT. THE FRACTURE MAY BE DUE TO A TRAUMATIC EVENT OCCURRED DURING THE WALKING RE-EDUCATION PERIOD. NO REASON TO SUSPECT THAT THIS EVENT IS DUE TO A FAULTY DEVICE. BATCH REVIEW PERFORMED ON (B)(6) 2017. LOT 153126: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06 OCTOBER 2015. EXPIRATION DATE: 2020-09-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT AVAILABLE.

Description of Event or Problem · 1

REVISION DUE TO PERIPROSTHETIC FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
159938 AMISTEM H, HA COATED STD SIZE 4 STEM CEMENTLESS FEMORAL STEM LZO MEDACTA INTERNATIONAL SA 153126

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention