FDA Adverse Event Malfunction Summary report: N

COAGULATION FACTOR VIII DEFICIENT PLASMA

MDR report key: 6376100 · Received March 3, 2017

Report

Report Number
9610806-2017-00028
Event Type
Malfunction
Date Received
March 3, 2017
Date of Event
February 2, 2017
Report Date
March 3, 2017
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Product Code
GJT
PMA / PMN Number
K924396
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER QUESTIONING THE HIGHER RESULTS OBTAINED ON THE BCS XP RELATIVE TO OTHER NON-SIEMENS SYSTEMS. SIEMENS HEADQUARTERS SUPPORT CENTER (HSC) REPRESENTATIVES HAVE EVALUATED THE INFORMATION PROVIDED. HSC CONCLUDED THAT REVIEW OF ALL AVAILABLE INFORMATION GIVES NO INDICATION OF A SYSTEMIC ISSUE WITH THE BCS XP SYSTEM, REAGENT MATERIAL, OR LABORATORY PRACTICES AT THE CUSTOMER LABORATORY. ROUTINE QUALITY CONTROL AS WELL AS PRIOR SURVEY SAMPLES RECOVER ACCEPTABLY. AN OVERALL PROBLEM WITH THE FVIII % ACTIVITY APPLICATION WITH ACTIN FSL ON THE BCS XP SYSTEM CAN BE EXCLUDED. PATIENT COMPARISONS HAVE BEEN CONDUCTED AND DEMONSTRATE ACCEPTABLE COMPARISON FOR FVIII % ACTIVITY BETWEEN THE ALTERNATE VENDOR METHODS (NON-SIEMENS). RECOVERY OF THE SAMPLES FVIII % ACTIVITY ON AN ALTERNATE INSTRUMENT SYSTEM WITH ACTIN FSL REAGENT AND FACTOR VIII DEFICIENT PLASMA ALSO WAS CONSISTENT WITH THE BCS XP SYSTEM RECOVERY. ADDITIONALLY, COMPARISON ACROSS VARIOUS LOTS OF ACTIN FSL COMPARE WELL. COMPARISON OF MULTIPLE EXTERNAL QUALITY ASSESSMENT SURVEY VENDOR SAMPLES DO NOT DEMONSTRATE A SYSTEMIC BIAS. A CLEAR DIFFERENCE IN RECOVERY WITH THIS SPECIFIC SAMPLE IS DEMONSTRATED BETWEEN THE APPLICATION OF SIEMENS FVIII WITH ACTIN FSL ON THE BCSP XP AS COMPARED TO COMPETITORS SYSTEMS. THE PATIENT HAD RECEIVED HELIXATE FS PRODUCT (RECOMBINANT FACTOR VIII CONCENTRATE) PRIOR TO COLLECTION OF THIS SAMPLE. THE RECOVERY WITH SIEMENS PRODUCTS IS BELIEVED TO BE ACCURATE FOR THE SAMPLE PROVIDED TO THE LABORATORY FOR ANALYSIS. A FALSE LOW BIAS IS BELIEVED TO BE DEMONSTRATED WITH THE ALTERNATE VENDOR PRODUCTS. THE CUSTOMER HAS CONFIRMED THAT NO IRREPARABLE HARM WAS CAUSED TO THE PATIENT BASED UPON THE ORIGINAL VALUE REPORTED. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FACTOR VIII (FVIII) RESULT ON A PATIENT SAMPLE WAS QUESTIONED AS DISCORDANT RELATIVE TO VALUES OBTAINED WITH ALTERNATE REAGENT METHODOLOGIES. THE SAMPLE WAS TESTED WITH COAGULATION FACTOR VIII DEFICIENT PLASMA USING THE DADE ACTIN FSL REAGENT ON THE BCS XP INSTRUMENT. THE SAME SAMPLE WAS SENT TO ALTERNATE LABORATORIES FOR TESTING. LOWER RESULTS WERE OBTAINED WITH TWO ALTERNATE NON-SIEMENS METHODOLOGIES ON NON-SIEMENS INSTRUMENTS. THE PHYSICIAN QUESTIONED THE INITIAL FVIII RESULTS WITH THE SIEMENS SYSTEM AS DISCORDANT WITH THE CLINICAL PRESENTATION. PATIENT DID RECEIVE FVIII CONCENTRATE FOUR HOURS BEFORE THE TESTING FOR FVIII WAS PERFORMED. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE QUESTIONED AS DISCORDANT FVIII RESULT. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE QUESTIONED AS DISCORDANT FVIII RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
160881 COAGULATION FACTOR VIII DEFICIENT PLASMA COAGULATION FACTOR VIII DEFICIENT PLASMA GJT SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH 547633

Patients

Seq Age Sex Outcome Treatment
1